Mechanical analysis of explanted silicone breast implants

被引:28
作者
Greenwald, DP
Randolph, M
May, JW
机构
[1] UNIV S FLORIDA,TAMPA,FL
[2] MASSACHUSETTS GEN HOSP,BOSTON,MA 02114
[3] HARVARD UNIV,SCH MED,BOSTON,MA
关键词
D O I
10.1097/00006534-199608000-00008
中图分类号
R61 [外科手术学];
学科分类号
摘要
Smooth-walled silicone gel breast implants (n = 25) were removed from 15 women after implantation times that varied from 23 to 216 months (mean = 117 months; SEM = 12.9). Strips of implant shells were tested to failure by computer-controlled tensiometer. Regression analysis revealed significant negative correlation between age of implantation and shell strength (p < 0.05), shell toughness (p < 0.05), and shell elasticity (p < 0.05). These data suggest that exposure to the in vivo environment weakens silicone gel breast implant shells over time.
引用
收藏
页码:269 / 272
页数:4
相关论文
共 8 条
[1]  
Fung YC., 1981, BIOMECHANICS
[2]   BIOMECHANICAL ANALYSIS OF INTRINSIC TENDON HEALING INVITRO AND THE EFFECTS OF VITAMIN-A AND VITAMIN-E [J].
GREENWALD, D ;
MASS, D ;
GOTTLIEB, L ;
TUEL, R .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1991, 87 (05) :925-930
[3]  
GREENWALD D, IN PRESS J SURG RES
[5]  
KOSSOVSKY N, 1983, PLAST RECONSTR SURG, V71, P795, DOI 10.1097/00006534-198306000-00008
[6]   ANALYSIS OF EXPLANTED SILICONE IMPLANTS - A REPORT OF 300 PATIENTS [J].
ROBINSON, OG ;
BRADLEY, EL ;
WILSON, DS .
ANNALS OF PLASTIC SURGERY, 1995, 34 (01) :1-6
[7]  
Vincent J., 2012, STRUCTURAL BIOMATERI, DOI [10.1515/9781400842780, DOI 10.1515/9781400842780]
[8]  
WAINWRIGHT SA, 1976, MECHANICAL DESIGN OR