Accuracy of Bioelectrical Impedance Consumer Devices for Measurement of Body Composition in Comparison to Whole Body Magnetic Resonance Imaging and Dual X-Ray Absorptiometry

被引:190
作者
Bosy-Westphal, Anja [1 ]
Later, Wiebke [1 ]
Hitze, Britta [1 ]
Sato, Tetsuya [2 ]
Kossel, Elke [3 ]
Glueer, Claus-Christian [3 ]
Heller, Martin [3 ]
Mueller, Manfred James [1 ]
机构
[1] Univ Kiel, Inst Humanernahrung & Lebensmittelkunde, D-24105 Kiel, Germany
[2] Omron Healthcare Co Ltd, Kyoto, Japan
[3] Univ Klinikum Schleswig Holstein, Radiol Diagnost Klin, Kiel, Germany
关键词
Body composition; Bioelectrical impedance analysis; Accuracy; Magnetic resonance imaging; Dual-energy X-ray absorptiometry;
D O I
10.1159/000176061
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To compare body composition determined by bioelectrical impedance (BIA) consumer devices against criterion estimates determined by whole body magnetic resonance imaging (MRI) and dual energy X-ray absorptiometry (DXA) in healthy normal weight, overweight and obese adults. Methods: In 106 adults (54 females, 52 males, age 54.2 +/- 16.1 years, BMI 25.8 +/- 4.4 kg/m(2)) fat mass (FM), skeletal muscle mass (SM), total body bone-free lean mass (TBBLM), and level of visceral fat mass (VF) were estimated by 3 single-frequency bipedal (foot-to-foot) and one tretrapolar BIA device, and compared to body composition measured by MRI and DXA. Bland-Altman and simple linear regression analyses were used to determine agreement between methods. Results: %FMDXA, SMMRI or TBBLMDXA showed good relative and absolute agreement with two bipolar and one tetrapolar instrument (r(2) = 0.92-0.96; all p < 0.001; mean bias <1.5 %FM and <1 kg SM or TBBLM) and less relative and absolute agreement for another bipolar device W = 0.82 and 0.84, mean bias similar to 3 %FM and similar to 3 kg SM). The 95% limits of agreement (bias +/- 2 SD) were narrowest for the tetrapolar device (-6.59 to 4.61 %FM and -4.62 to 4.74 kg SM) and widest for bipolar instruments (up to -14.54 to 8.58 %FM and -9.52 to 3.92 kg SM). Systematic biases for %FM were found for all bipedal devices, but not for the tetrapolar instrument. Conclusion: Because of the lower agreement between foot-to-foot BIA and DXA or MRI for the assessment of body composition in individuals, tetrapolar electrode arrangement should be preferred for individual or public use. Bipolar devices provide accurate results for field studies with group estimation.
引用
收藏
页码:319 / 324
页数:6
相关论文
共 29 条
[1]  
Andreoli A, 2002, J SPORT MED PHYS FIT, V42, P186
[2]   Intra- and interindividual variability of resting energy expenditure in healthy male subjects -: biological and methodological variability of resting energy expenditure [J].
Bader, N ;
Bosy-Westphal, A ;
Dilba, B ;
Müller, MJ .
BRITISH JOURNAL OF NUTRITION, 2005, 94 (05) :843-849
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
Buchholz Andrea C, 2004, Nutr Clin Pract, V19, P433, DOI 10.1177/0115426504019005433
[5]   A whole-body model to distinguish excess fluid from the hydration of major body tissues [J].
Chamney, Paul W. ;
Wabel, Peter ;
Moissl, Ulrich M. ;
Mueller, Manfred J. ;
Bosy-Westphal, Anja ;
Korth, Oliver ;
Fuller, Nigel J. .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 2007, 85 (01) :80-89
[6]  
Cornish BH, 1996, EUR J CLIN NUTR, V50, P159
[7]  
Demura Shinichi, 2004, Journal of Physiological Anthropology and Applied Human Science, V23, P93, DOI 10.2114/jpa.23.93
[8]  
Deurenberg Paul, 1996, American Journal of Clinical Nutrition, V64, p449S, DOI 10.1093/ajcn/64.3.449S
[9]   Is a re-evaluation of WHO body mass index cut-off values needed? The case of Asians in Singapore [J].
Deurenberg-Yap, M ;
Deurenberg, P .
NUTRITION REVIEWS, 2003, 61 (05) :S80-S87
[10]   Selected body composition methods can be used in field studies [J].
Ellis, KJ .
JOURNAL OF NUTRITION, 2001, 131 (05) :1589S-1595S