Identifying and tracking a guide wire in the coronary arteries during angioplasty from X-ray images

被引:39
作者
PaltiWasserman, D [1 ]
Brukstein, AM [1 ]
Beyar, RP [1 ]
机构
[1] TECHNION ISRAEL INST TECHNOL, JULIUS SILVER INST, FAC COMP SCI, IL-32000 HAIFA, ISRAEL
关键词
coronary angioplasty; guidewire; tracking; ventricular function; wall motion;
D O I
10.1109/10.552245
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
During angioplasty, a guide mire (GW) is routinely placed in the coronary artery, Balloon inflation during angioplasty causes transient occlusion of the coronary artery and regional dysfunction. Thus, it is of major importance to monitor myocardial function, which may be impaired during this period. Since the GW moves with the coronary arteries, information regarding myocardial function can potentially be extracted from the GW motion. An algorithm is suggested which is a step toward such monitoring, The algorithm presented is a semiautomatic procedure for identifying and tracking the GW using specific characteristics of the GW, This algorithm is based on working in limited active windows. A preprocessing stage which enhances the GW by the use of a modified Laplacian filter or a modified Marr-Hildreth filter is introduced, The second stage of the algorithm is the tracking of the GW, which is based on fitting a second-degree polynom to the GW using the Dough transform in each window. To further improve the results further modifications of the basic algorithms that were taken, A single set of parameters, which enabled good tracking for a large number of images taken during angioplasty, was fitted to the final algorithm.
引用
收藏
页码:152 / 164
页数:13
相关论文
共 37 条
[1]   REGIONAL HEART WALL MOTION - 2-DIMENSIONAL ANALYSIS AND FUNCTIONAL IMAGING WITH MR IMAGING [J].
AXEL, L ;
GONCALVES, RC ;
BLOOMGARDEN, D .
RADIOLOGY, 1992, 183 (03) :745-750
[2]   3-DIMENSIONAL MAPPING OF ACUTE ISCHEMIC REGIONS USING MRI - WALL THICKENING VERSUS MOTION ANALYSIS [J].
AZHARI, H ;
SIDEMAN, S ;
WEISS, JL ;
SHAPIRO, EP ;
WEISFELDT, ML ;
GRAVES, WL ;
ROGERS, WJ ;
BEYAR, R .
AMERICAN JOURNAL OF PHYSIOLOGY, 1990, 259 (05) :H1492-H1503
[3]  
Azuma A, 1989, J Cardiol, V19, P351
[4]   GENERALIZING THE HOUGH TRANSFORM TO DETECT ARBITRARY SHAPES [J].
BALLARD, DH .
PATTERN RECOGNITION, 1981, 13 (02) :111-122
[5]   LEFT-VENTRICULAR SYSTOLIC AND DIASTOLIC FUNCTION DURING ACUTE CORONARY-ARTERY BALLOON OCCLUSION IN HUMANS [J].
BERTRAND, ME ;
LABLANCHE, JM ;
FOURRIER, JL ;
TRAISNEL, G ;
MIRSKY, I .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 12 (02) :341-347
[6]  
BEYAR R, 1990, IMAGING ANAL SIMULAT, P137
[7]  
BHARGAVA V, 1990, COMPUT MED IMAG GRAP, V14, P107
[8]   QUANTITATIVE ASSESSMENT OF TEMPORAL AND SPATIAL VENTRICULAR WALL MOTION IN NORMAL AND INFARCTED HUMAN LEFT-VENTRICLES [J].
CHEN, C ;
BONZEL, T ;
JUST, H ;
SEIFFERT, G ;
ZEIHER, A ;
KASPER, W ;
WOLLSCHLAEGER, H .
AMERICAN HEART JOURNAL, 1986, 112 (04) :712-721
[9]  
CHEN C, 1988, SPIE, V914, P594
[10]  
DEFFE I, 1989, P 6 SCAND C IM AN OU, P1134