A simple, reliable, reproducible and sensitive method, based on capillary electrophoresis with electrochemical detection (ED), for the determination of liquiritigenin and isoliquiritigenin in Glycyrrhiza uralensis and its medicinal preparations was described. Operated in a wall-jet configuration, a 300 mum diameter carbon-disk electrode was used as the working electrode, which exhibits good responses at +1000 mV (versus SCE) for the two analytes. Under the optimum conditions, the analytes were base-line separated within 8 min, and excellent linearity was obtained in the concentration range from 5.0 X 10(-4) to 1.0 X 10(-6) mol/l. The detection limit (S/N = 3) was 4.7 x 10(-7) and 2.9 X 10(-7) Mol/l for liquiritigenin and isoliquiritigenin, respectively. This work provides a useful method for the analysis of traditional Chinese medicines. (C) 2004 Elsevier B.V. All rights reserved.