Flow-through amperometric immunosensor: Fast 'sandwich' scheme immunoassay

被引:44
作者
Ghindilis, AL [1 ]
Krishnan, R [1 ]
Atanasov, P [1 ]
Wilkins, E [1 ]
机构
[1] UNIV NEW MEXICO,DEPT CHEM & NUCL ENGN,ALBUQUERQUE,NM 87131
关键词
amperometric immunosensor; immunoelectrode; flow-through immunoassay;
D O I
10.1016/S0956-5663(97)00016-X
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A flow-through immunosensor based on a high-surface-area carbon immunoelectrode has been developed. Dispersed carbon material serves as a carrier for immobilized antibodies and at the same time as an electrode material. The 'sandwich' scheme of immunoassay has been used. Iodine formed as a result of the enzymatic oxidation of iodide by a peroxidase label has been detected amperometrically. The immunosensor consists of a disposable sensing element (immunocolumn) containing dispersed carbon material with immobilized antibodies which also acts as a working electrode. A current collector connects the working electrode to the measuring device. The electrochemical detection time of the peroxidase-labeled immune-complex does not exceed several minutes. The overall time of analysis including flowing of analyte, flowing of antigen, washing and detection stages is as low as 22min. This technique allows fast determination of rabbit IgG (used as a model analyte) with a low detection limit in the picomolar range and also the determination of human IgM in blood plasma with a low detection limit in the nanomolar range. (C) 1997 Elsevier Science Limited.
引用
收藏
页码:415 / 423
页数:9
相关论文
共 26 条
  • [1] Zeptomole-detecting biosensor for alkaline phosphatase in an electrochemical immunoassay for 2,4-dichlorophenoxyacetic acid
    Bauer, CG
    Eremenko, AV
    EhrentreichForster, E
    Bier, FF
    Makower, A
    Halsall, HB
    Heineman, WR
    Scheller, FW
    [J]. ANALYTICAL CHEMISTRY, 1996, 68 (15) : 2453 - 2458
  • [2] IMMUNOBASED ELUTION ASSAY FOR PROCESS-CONTROL
    BEYER, K
    REINECKE, M
    NOE, W
    SCHEPER, T
    [J]. ANALYTICA CHIMICA ACTA, 1995, 309 (1-3) : 301 - 305
  • [3] BUCKLEY E, 1989, Analytical Proceedings, V26, P5
  • [4] MEASUREMENT OF ALKALINE-PHOSPHATASE ACTIVITY BY ELECTROCHEMICAL DETECTION OF PHOSPHATE-ESTERS - APPLICATION TO AMPEROMETRIC ENZYME-IMMUNOASSAY
    FREW, JE
    FOULDS, NC
    WILSHERE, JM
    FORROW, NJ
    GREEN, MJ
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1989, 266 (02): : 309 - 316
  • [5] DETERMINATION OF P-AMINOPHENOL AND CATECHOLAMINES AT PICOMOLAR CONCENTRATIONS BASED ON RECYCLING ENZYME AMPLIFICATION
    GHINDILIS, AL
    MAKOWER, A
    BAUER, CG
    BIER, FF
    SCHELLER, FW
    [J]. ANALYTICA CHIMICA ACTA, 1995, 304 (01) : 25 - 31
  • [6] GHINDILIS AL, 1995, ANAL LETT, V28, P1
  • [7] SIMULTANEOUS IMMUNOASSAY USING ELECTROCHEMICAL DETECTION OF METAL-ION LABELS
    HAYES, FJ
    HALSALL, HB
    HEINEMAN, WR
    [J]. ANALYTICAL CHEMISTRY, 1994, 66 (11) : 1860 - 1865
  • [8] A DISPOSABLE AMPEROMETRIC IMMUNOSENSOR FOR 2,4-DICHLOROPHENOXYACETIC ACID
    KALAB, T
    SKLADAL, P
    [J]. ANALYTICA CHIMICA ACTA, 1995, 304 (03) : 361 - 368
  • [9] ELECTROCHEMICAL ENZYME-IMMUNOASSAY USING SEQUENTIAL SATURATION TECHNIQUE IN A 20-MUL CAPILLARY - DIGOXIN AS A MODEL ANALYTE
    KANEKI, N
    XU, Y
    KUMARI, A
    HALSALL, HB
    HEINEMAN, WR
    KISSINGER, PT
    [J]. ANALYTICA CHIMICA ACTA, 1994, 287 (03) : 253 - 258
  • [10] Krishnan R, 1996, ANAL LETT, V29, P2615