The Cleveland clinic rotodynamic pump program

被引:11
作者
Golding, LAR
Smith, WA
Bodmann, DR
机构
关键词
rotodynamic; blood pump; ventricular assist system;
D O I
10.1111/j.1525-1594.1996.tb04468.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Cleveland Clinic Foundation has developed a unique rotodynamic blood pump for future use as a permanent implant. This pump is small (2.5 x 2.5 inches) and requires an electric input power of 7 watts to produce 5 L/min of blood flow against 100 mm Hg at 3,000 rpm. Initial in vivo testing has confirmed in vitro function and shown low hemolysis. Endurance bench testing has exceeded 12 months of continuous function. This pump is the basis of an innovative ventricular assist system in which power is supplied by a tranocutaneous electrical transmission system, and pacer technology is used for both control logic and telemetry functions. The resulting system will be completed and tested under an NHLBI contract during the next 5 years.
引用
收藏
页码:481 / 484
页数:4
相关论文
共 8 条
[1]  
GOLDING I, 1994, Patent No. 5324177
[2]   POSTCARDIOTOMY CENTRIFUGAL MECHANICAL VENTRICULAR SUPPORT [J].
GOLDING, LAR ;
CROUCH, RD ;
STEWART, RW ;
NOVOA, R ;
LYTLE, BW ;
MCCARTHY, PM ;
TAYLOR, PC ;
LOOP, FD ;
COSGROVE, DM .
ANNALS OF THORACIC SURGERY, 1992, 54 (06) :1059-1064
[3]   RESULTS OF MECHANICAL VENTRICULAR ASSIST IN BRIDGING TO CARDIAC TRANSPLANTATION [J].
GOLDING, LAR ;
TISHKO, DJ ;
STEWART, RW .
CLEVELAND CLINIC JOURNAL OF MEDICINE, 1988, 55 (01) :59-62
[4]  
GOLDING LAR, 1991, Patent No. 5049134
[5]  
GOLDING LAR, 1994, Patent No. 5370509
[6]  
GOLDING LR, 1980, T AM SOC ART INT ORG, V26, P251
[7]  
GOLDING LR, 1978, T AM SOC ART INT ORG, V24, P93
[8]  
YADA I, 1983, T AM SOC ART INT ORG, V29, P520