An MRI-compatible system for focused ultrasound experiments in small animal models

被引:84
作者
Chopra, Rajiv [1 ,2 ]
Curiel, Laura [1 ,3 ]
Staruch, Robert [1 ,2 ]
Morrison, Laetitia [1 ]
Hynynen, Kullervo [1 ,2 ]
机构
[1] Sunnybrook Hlth Sci Ctr, Toronto, ON M4N 3M5, Canada
[2] Univ Toronto, Dept Med Biophys, Toronto, ON M5G 2M9, Canada
[3] Thunder Bay Reg Res Inst, HIFU Lab, Thunder Bay, ON P7B 6V4, Canada
基金
美国国家卫生研究院;
关键词
biomedical MRI; blood; brain; diseases; drug delivery systems; feedback; haemodynamics; hyperthermia; medical computing; medical control systems; neurophysiology; piezoelectric actuators; position control; ultrasonic focusing; ultrasonic therapy; BLOOD-BRAIN-BARRIER; MAGNETIC-RESONANCE; UTERINE LEIOMYOMAS; TARGETED DELIVERY; DISRUPTION; SURGERY; FEASIBILITY; INTERVENTIONS; RABBITS; BREAST;
D O I
10.1118/1.3115680
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The development of novel MRI-guided therapeutic ultrasound methods including potentiated drug delivery and targeted thermal ablation requires extensive testing in small animals such as rats and mice due to the widespread use of these species as models of disease. An MRI-compatible, computer-controlled three-axis positioning system was constructed to deliver focused ultrasound exposures precisely to a target anatomy in small animals for high-throughput preclinical drug delivery studies. Each axis was constructed from custom-made nonmagnetic linear ball stages driven by piezoelectric actuators and optical encoders. A range of motion of 5x5x2.5 cm(3) was achieved, and initial bench top characterization demonstrated the ability to deliver ultrasound to the brain with a spatial accuracy of 0.3 mm. Operation of the positioning system within the bore of a clinical 3 T MR imager was feasible, and simultaneous motion and MR imaging did not result in any mutual interference. The system was evaluated in its ability to deliver precise sonications within the mouse brain, linear scanned exposures in a rat brain for blood barrier disruption, and circular scans for controlled heating under MR temperature feedback. Initial results suggest that this is a robust and precise apparatus for use in the investigation of novel ultrasound-based therapeutic strategies in small animal preclinical models.
引用
收藏
页码:1867 / 1874
页数:8
相关论文
共 30 条
[1]   Control of thermal therapies with moving power deposition field [J].
Arora, D ;
Minor, MA ;
Skliar, M ;
Roemer, RB .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (05) :1201-1219
[2]  
Chinzei K, 2001, Med Sci Monit, V7, P153
[3]   Spatio-temporal analysis of molecular delivery through the blood-brain barrier using focused ultrasound [J].
Choi, J. J. ;
Pernot, M. ;
Brown, T. R. ;
Small, S. A. ;
Konofagou, E. E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (18) :5509-5530
[4]   FOCUSED US SYSTEM FOR MR IMAGING-GUIDED TUMOR ABLATION [J].
CLINE, HE ;
HYNYNEN, K ;
WATKINS, RD ;
ADAMS, WJ ;
SCHENCK, JF ;
ETTINGER, RH ;
FREUND, WR ;
VETRO, JP ;
JOLESZ, FA .
RADIOLOGY, 1995, 194 (03) :731-737
[5]   Progress in multimodality imaging: Truly simultaneous ultrasound and magnetic resonance imaging [J].
Curiel, Laura ;
Chopra, Rajiv ;
Hynyne, Kullervo .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2007, 26 (12) :1740-1746
[6]   The role of ultrasound and magnetic resonance in local drug delivery [J].
Deckers, Roel ;
Rome, Claire ;
Moonen, Chrit T. W. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (02) :400-409
[7]   MRI-guided focused ultrasound surgery of uterine leiomyomas [J].
Fennessy, FM ;
Tempany, CM .
ACADEMIC RADIOLOGY, 2005, 12 (09) :1158-1166
[8]   Ultrasound mediated delivery of drugs and genes to solid tumors [J].
Frenkel, Victor .
ADVANCED DRUG DELIVERY REVIEWS, 2008, 60 (10) :1193-1208
[9]  
Hempel Eckhard, 2003, Comput Aided Surg, V8, P180, DOI 10.3109/10929080309146052
[10]  
Huber PE, 2001, CANCER RES, V61, P8441