Efficacy of a filter device in the prevention of embolic events during carotid angioplasty and stenting: An ex vivo analysis

被引:151
作者
Ohki, T
Roubin, GS
Veith, FJ
Iyer, SS
Brady, E
机构
[1] Montefiore Med Ctr, Div Vasc Surg, Albert Einstein Coll Med, Dept Surg, Bronx, NY 10467 USA
[2] Lenox Hill Hosp, Lenox hill Heart & Vasc Inst, New York, NY 10021 USA
[3] Med Nova Inc, New York, NY USA
关键词
D O I
10.1016/S0741-5214(99)70041-8
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: Although percutaneous angioplasty and stenting (PTAS) of carotid bifurcation lesions is feasible and appropriate for surgically inaccessible lesions, its general role and comparative value remain unclarified. Moreover, the acceptance of carotid PTAS has been limited by its potential for producing embolic debris. This study used an ex vivo model to evaluate the efficacy of a novel filter device to entrap emboli during PTAS of human carotid plaques. Methods: Eight carotid bifurcation plaques were obtained from patients who underwent carotid endarterectomy for high-grade atherosclerotic stenosis(>90%). The mean age of the patients was 63 years, and six patients were symptomatic. Each plaque was encased with polytetrafluoroethylene material to simulate adventitia and was connected to a perfusion circuit, which provided continuous flow through the plaque. The filter device consisted of an expandable polymeric membrane with multiple micro pores that was attached to the distal end of a 0.014-in wire with a shapeable tip. This filter was encased in a delivery catheter. With fluoroscopic guidance, the filter wire was passed through the stenosis and the delivery catheter was then retracted to open the filter to capture particles released into the distal internal carotid artery. PTAS with a self-expandable stent then was carried out over the filter wire. The particles released during the initial filter passage, those captured in the filter, and those that flowed through or around the filter (missed) mere collected and analyzed with light microscopy. Results: Eater deployment, PTAS, and filter retrieval were achieved successfully with each lesion. Because the filter has a low crossing profile, it passed through the stenoses smoothly and only produced occasional small particles. PTAS improved the angiographic stenosis from 96.2% +/- 3.7% to 1.3% +/- 1.6%. The mean number and the maximum size of the particles that were released during initial filter passage, missed, and captured by the filter device were 3.1 and 500 mu m, 2.8 and 360 mu m, 20.1 and 1100 mu m, respectively. Most of the particles and those of large size were released during PTAS, The filter captured 88% of these particles. Conclusion: These study results show that this filter device, at least in this model, did not add complexity to the interventional procedure itself. Furthermore, the filter may markedly decrease embolic events during carotid PTAS and expand the indications for this procedure.
引用
收藏
页码:1034 / 1042
页数:9
相关论文
共 34 条
  • [1] THE SIGNIFICANCE OF MICROEMBOLI DETECTION BY MEANS OF TRANSCRANIAL DOPPLER ULTRASONOGRAPHY MONITORING IN CAROTID ENDARTERECTOMY
    ACKERSTAFF, RGA
    JANSEN, C
    MOLL, FL
    VERMEULEN, FEE
    HAMERLIJNCK, RPHM
    MAUSER, HW
    [J]. JOURNAL OF VASCULAR SURGERY, 1995, 21 (06) : 963 - 969
  • [2] Concern about safety of carotid angioplasty
    Beebe, HG
    Archie, JP
    Baker, WH
    Barnes, RW
    Becker, GJ
    Bernstein, EF
    Brener, B
    Clagett, GP
    Clowes, AW
    Cooke, JP
    Creager, MA
    Cronenwett, JL
    Dake, M
    DeWeese, JA
    Fogarty, TJ
    Freischlag, JA
    Goldstone, J
    Greenfield, LJ
    Hertzer, NR
    Hobson, RW
    Joyce, JW
    Katzen, BT
    LoGerfo, FW
    Mohr, JP
    Moore, WS
    Najafi, H
    Ricotta, JJ
    Riles, TS
    Ring, EJ
    Robertson, J
    Rutherford, RB
    Sos, T
    Stanley, JC
    Strandness, DE
    Sumner, DS
    Toole, J
    Towne, JB
    Veith, FJ
    Whittemore, AD
    Yao, JST
    Zarins, CK
    [J]. STROKE, 1996, 27 (02) : 197 - 198
  • [3] Bergeron P, 1996, J CARDIOVASC SURG, V37, P73
  • [4] Carotid stenting and angioplasty - A statement for healthcare professionals from the councils on cardiovascular radiology, stroke, cardio-thoracic and vascular surgery, epidemiology and prevention, and clinical cardiology, American heart association
    Bettmann, MA
    Katzen, BT
    Whisnant, J
    Brant-Zawadzki, M
    Broderick, JP
    Furlan, AJ
    Hershey, LA
    Howard, V
    Kuntz, R
    Loftus, CM
    Pearce, W
    Roberts, A
    Roubin, G
    [J]. STROKE, 1998, 29 (01) : 336 - 338
  • [5] Calvey TAJ, 1998, J VASC SURG, V27, P753
  • [6] Evolving indications for and early results of carotid artery stenting
    Criado, FJ
    Wellons, E
    Clark, NS
    [J]. AMERICAN JOURNAL OF SURGERY, 1997, 174 (02) : 111 - 114
  • [7] CAROTID TRANS-LUMINAL ANGIOPLASTY WITH EVIDENCE OF DISTAL EMBOLIZATION - CASE-REPORT
    DEMONTE, F
    PEERLESS, SJ
    RANKIN, RN
    [J]. JOURNAL OF NEUROSURGERY, 1989, 70 (01) : 138 - 141
  • [8] Stent-supported carotid angioplasty - Should it be done, and, if so, by whom? A 1998 perspective
    Dorros, G
    [J]. CIRCULATION, 1998, 98 (09) : 927 - 930
  • [9] CLINICAL RELEVANCE OF INTRAOPERATIVE EMBOLIZATION DETECTED BY TRANSCRANIAL DOPPLER ULTRASONOGRAPHY DURING CAROTID ENDARTERECTOMY - A PROSPECTIVE-STUDY OF 100 PATIENTS
    GAUNT, ME
    MARTIN, PJ
    SMITH, JL
    RIMMER, T
    CHERRYMAN, G
    RATLIFF, DA
    BELL, PRF
    NAYLOR, AR
    [J]. BRITISH JOURNAL OF SURGERY, 1994, 81 (10) : 1435 - 1439
  • [10] Henry M, 1998, J ENDOVASC SURG, V5, P293, DOI 10.1583/1074-6218(1998)005<0293:AASOTE>2.0.CO