Comparison of realizations of the thermometric fixed points of the ITS-90 in the range -38 degrees C to 420 degrees C between Canada and Mexico

被引:4
作者
Ancsin, J
MendezLango, E
机构
[1] Inst. for Natl. Msrmt. Standards, National Research Council of Canada, Ottawa
[2] Centro Nacional de Metrología, Division Termometría, Municipio de El Marques Queretaro, CP 76900 QRO
关键词
D O I
10.1088/0026-1394/33/4/18
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The realizations of the fixed points of the International Temperature Scale of 1990 (ITS-90) maintained in Canada and Mexico were compared between -38 degrees C and 420 degrees C, using a portable fixed-point realization apparatus. In this range, the routinely realized defining fixed points of the ITS-90 in these countries were found to coincide to within 0,5 mK.