The impact of vessel orientation in space on densitometric measurements of cross sectional areas of coronary arteries

被引:2
作者
Doriot, PA
Dorsaz, PA
Dorsaz, L
Chatelain, P
Rutishauser, W
机构
[1] Cardiology Center, University Hospital, Geneva
[2] Cardiology Center, University Hospital
来源
INTERNATIONAL JOURNAL OF CARDIAC IMAGING | 1996年 / 12卷 / 04期
关键词
coronary angiography; QCA; densitometry;
D O I
10.1007/BF01797742
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Under ideal conditions, densitometric measurement of a coronary arterial cross section in biplane angiographic images should result in nearly equal cross sectional areas for both planes. However, quite appreciable discrepancies have been found by some authors in patients. In this study, the role of inadequate spatial orientation of the vessel axes relatively to the x-rays was assessed by use of a 3D technique applied to 60 stenoses (45 pre PTCA and 15 post PTCA) in simultaneously acquired digital biplane coronary angiograms of 27 CAD patients. The 3D technique yields two radius values per projection directly in mm at any arterial cross section of interest. This was used to determine the areas A(r)(in mm(2)) of the reference cross sections. As with catheter calibration, these cross sections were thus assumed to be more or less circular, but out-of-plane effects and errors due to a catheter diameter determination in pixels were avoided. The areas of the stenotic sections were then determined densitometrically (in mm(2)) from the two projections (1 and 2) according to A(s1) = A(r)D(s1)/D-r1, resp. A(s2) = A(r)D(s2)/D-r2, where D-r1, D-r2, D-s1 and D-s2 are the conventional densitometric areas of the reference and stenotic cross sections measured in planes 1 and 2. As expected, the areas A(s1) and A(s2) correlated only moderately: A(s2) = 0.92 A(s1) + 0.7 mm(2), r = 0.82, n = 60, SEE = 1.4 mm(2). The 3D method also yielded the two spatial angles between the local vessel axis and the x-rays of both planes. These two angles were then used to correct each densitometric area for inadequate orientation. With the corrected densitometric areas A(s1c) and A(s2c), the correlation improved to: A(s2c) = 1.05 A(s1c) + 0.03 mm(2), r = 0.93, n = 60, SEE = 0.8 mm(2). Inadequate orientation of the cross sections in space thus appears to be an important factor of inaccuracy in densitometric measurements of stenotic cross sections in patients.
引用
收藏
页码:289 / 297
页数:9
相关论文
共 28 条
[1]  
Balkin J, 1990, Int J Card Imaging, V5, P119, DOI 10.1007/BF01833980
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]  
DIMARIO C, 1992, AM HEART J, V124, P1181
[4]  
Doriot P.-A., 1982, Computers in Cardiology, P329
[5]  
Doriot P A, 1992, Int J Card Imaging, V8, P121, DOI 10.1007/BF01137533
[6]   MORPHOMETRIC VERSUS DENSITOMETRIC ASSESSMENT OF CORONARY VASOMOTOR TONE - AN OVERVIEW [J].
DORIOT, PA ;
GUGGENHEIM, N ;
DORSAZ, PA ;
RUTISHAUSER, W .
EUROPEAN HEART JOURNAL, 1989, 10 :49-53
[7]  
DORIOT PA, 1986, COMPUT CARDIOL, P91
[8]   QUANTITATIVE ANGIOGRAPHY DURING CORONARY ANGIOPLASTY WITH A SINGLE ANGIOGRAPHIC VIEW - A COMPARISON OF AUTOMATED EDGE-DETECTION AND VIDEODENSITOMETRIC TECHNIQUES [J].
ESCANED, J ;
FOLEY, DP ;
HAASE, J ;
DIMARIO, C ;
HERMANS, WR ;
DEFEYTER, PJ ;
SERRUYS, PW .
AMERICAN HEART JOURNAL, 1993, 126 (06) :1326-1333
[9]   PITFALLS IN THE DETERMINATION OF ABSOLUTE DIMENSIONS USING ANGIOGRAPHIC CATHETERS AS CALIBRATION DEVICES IN QUANTITATIVE ANGIOGRAPHY [J].
FORTIN, DF ;
SPERO, LA ;
CUSMA, JT ;
SANTORO, L ;
BURGESS, R ;
BASHORE, TM .
AMERICAN JOURNAL OF CARDIOLOGY, 1991, 68 (11) :1176-1182
[10]   SPATIAL RECONSTRUCTION OF CORONARY-ARTERIES FROM ANGIOGRAPHIC IMAGES [J].
GUGGENHEIM, N ;
DORIOT, PA ;
DORSAZ, PA ;
DESCOUTS, P ;
RUTISHAUSER, W .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (01) :99-110