Determination of Ir and Pt down to the sub-femtogram per gram level in polar ice by ICP-SFMS using preconcentration and a desolvation system

被引:31
作者
Gabrielli, P
Varga, A
Barbante, C [1 ]
Boutron, C
Cozzi, G
Gaspari, V
Planchon, F
Cairns, W
Hong, SM
Ferrari, C
Capodaglio, G
机构
[1] Univ Venice, Dept Environm Sci, I-30123 Venice, Italy
[2] Univ Venice, Dept Environm Sci, I-30123 Venice, Italy
[3] Univ Venice, CNR, Inst Dynam Environm Proc, I-30123 Venice, Italy
[4] Univ Grenoble 1, Observ Sci Univ, F-38041 Grenoble, France
[5] Univ Grenoble 1, Unite Format & Rech Phys, Inst Univ France, F-38041 Grenoble, France
[6] Korea Polar Res Inst, Korea Ocean Res & Dev Inst, Seoul 425600, South Korea
[7] Univ Grenoble 1, Ecole Polytech Univ Grenoble, Inst Univ France, F-38041 Grenoble, France
[8] CNRS, Lab Glaciol & Geophys Environm, F-38402 St Martin Dheres, France
关键词
D O I
10.1039/b316283d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new analytical methodology, based on inductively coupled plasma sector field mass spectrometry (ICP-SFMS) coupled with a micro-flow nebulizer and desolvation system, has been set up for the quantification of Ir and Pt down to the sub-ppq level (1 ppq=1 fg g(-1)=10(-15) g g(-1)) in polar ice samples. Ultra-clean procedures were adopted during the pre-treatment phases in our laboratories in order to avoid possible contamination problems and a preconcentration step by evaporation at sub-boiling temperatures was necessary. A procedural detection limit of 0.02 ppq and 0.08 ppq for Ir and Pt, respectively, was obtained. The reproducibility of the analytical procedure at the ppq level was about 50% for Ir and 30% for Pt and the recoveries were 75% and 93% for Ir and Pt, respectively. Spectral interferences, which affect the determination of Ir and Pt, were reduced by using a desolvation system for sample introduction. The contribution of the interfering species was determined and subtracted. This new method allowed us to analyse Ir and Pt in remote uncontaminated ice samples from Antarctica and Greenland down to the sub-ppq level. The concentration ranges were from 0.1 up to 5 ppq for Ir and from 0.2 up to 7 ppq for Pt. These measurements represent the first data of Ir concentrations in unfiltered melted ice samples and the lowest concentrations ever recorded for Pt in environmental samples.
引用
收藏
页码:831 / 837
页数:7
相关论文
共 35 条
[1]   EXTRATERRESTRIAL CAUSE FOR THE CRETACEOUS-TERTIARY EXTINCTION - EXPERIMENTAL RESULTS AND THEORETICAL INTERPRETATION [J].
ALVAREZ, LW ;
ALVAREZ, W ;
ASARO, F ;
MICHEL, HV .
SCIENCE, 1980, 208 (4448) :1095-1108
[2]   Iridium in natural waters [J].
Anbar, AD ;
Wasserburg, GJ ;
Papanastassiou, DA ;
Andersson, PS .
SCIENCE, 1996, 273 (5281) :1524-1528
[3]   Determination of iridium in natural waters by clean chemical extraction and negative thermal ionization mass spectrometry [J].
Anbar, AD ;
Papanastassiou, DA ;
Wasserburg, GJ .
ANALYTICAL CHEMISTRY, 1997, 69 (13) :2444-2450
[4]   ABUNDANCES OF THE ELEMENTS - METEORITIC AND SOLAR [J].
ANDERS, E ;
GREVESSE, N .
GEOCHIMICA ET COSMOCHIMICA ACTA, 1989, 53 (01) :197-214
[5]   Determination of Rh, Pd, and Pt in polar and alpine snow and ice by double focusing ICPMS with microconcentric nebulization [J].
Barbante, C ;
Cozzi, G ;
Capodaglio, G ;
Van de Velde, K ;
Ferrari, C ;
Veysseyre, A ;
Boutron, CF ;
Scarponi, G ;
Cescon, P .
ANALYTICAL CHEMISTRY, 1999, 71 (19) :4125-4133
[6]   Greenland snow evidence of large scale atmospheric contamination for platinum, palladium, and rhodium [J].
Barbante, C ;
Veysseyre, A ;
Ferrari, C ;
Van de Velde, K ;
Morel, C ;
Capodaglio, G ;
Cescon, P ;
Scarponi, G ;
Boutron, C .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2001, 35 (05) :835-839
[7]   Direct determination of heavy metals at picogram per gram levels in Greenland and Antarctic snow by double focusing inductively coupled plasma mass spectrometry [J].
Barbante, C ;
Bellomi, T ;
Mezzadri, G ;
Cescon, P ;
Scarponi, G ;
Morel, C ;
Jay, S ;
VanDeVelde, K ;
Ferrari, C ;
Boutron, CF .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (09) :925-931
[8]   PRELIMINARY DATA ON CHANGES OF LEAD CONCENTRATIONS IN ANTARCTIC ICE FROM 155,000 TO 26,000 YEARS BP [J].
BOUTRON, CF ;
PATTERSON, CC ;
PETROV, VN ;
BARKOV, NI .
ATMOSPHERIC ENVIRONMENT, 1987, 21 (05) :1197-1202
[9]   A CLEAN LABORATORY FOR ULTRALOW CONCENTRATION HEAVY-METAL ANALYSIS [J].
BOUTRON, CF .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 337 (05) :482-491
[10]   LEAD CONCENTRATION CHANGES IN ANTARCTIC ICE DURING THE WISCONSIN HOLOCENE TRANSITION [J].
BOUTRON, CF ;
PATTERSON, CC .
NATURE, 1986, 323 (6085) :222-225