The use of a Hamamatsu X-ray image intensifier with a cooled CCD as a solution X-ray scattering detector

被引:48
作者
Fujisawa, T
Inoko, Y
Yagi, N
机构
[1] RIKEN, Harima Inst, SPring 8, Struct Biophys Lab, Sayo, Hyogo 6795148, Japan
[2] Osaka Univ, Grad Sch Engn Sci, Div Biophys Engn, Toyonaka, Osaka 5608531, Japan
[3] Japan Synchrotron Radiat Res Inst, JASRI, Expt Res Div, SPring 8, Sayo, Hyogo 6795198, Japan
关键词
SPring-8; solution X-ray scattering; X-ray image intensifier with cooled CCD; apoferritin;
D O I
10.1107/S0909049599010250
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
CCD detectors are now widely used in many synchrotron small-angle X-ray scattering beamlines. The use of an X-ray image intensifier with cooled CCD (XR-II + CCD) was studied, especially for use in synchrotron solution X-ray scattering. Two samples, polystyrene latex and apoferritin, were used. These two samples have fine structure in the solution scattering profile due to symmetry and narrow size distribution. The recorded scattering profile, in comparison with that obtained by a position-sensitive proportional counter (PSPC), showed that XR-II + CCD has a much smaller practical dynamic range (100:1) than that of a pixel well (7500:1). This limited dynamic range was overcome by placing various-size masks on the detection plane, thereby eliminating the high-intensity region. The images recorded with various masks were combined, and the reconstituted solution scattering profile was submitted to various analyses, including Guinier analysis, power-law analysis, size distribution analysis and calculation of radial density distributions. The results were the same as those obtained with the PSPC. This indicates that spatial distortion as well as shading, a decrease in sensitivity from the centre to the edge of the detecting region [Amemiya, Ito, Yagi, Asano, Wakabayashi, Ueki & Endo (1995). Rev. Sci. Instrum. 66, 2290-2294], have very little effect on the SBXS results. This paper presents a practical protocol for obtaining a reliable solution scattering profile given the limitations of XR-II + CCD for synchrotron solution X-ray scattering.
引用
收藏
页码:1106 / 1114
页数:9
相关论文
共 18 条
  • [1] LARGE-APERTURE TV DETECTOR WITH A BERYLLIUM-WINDOWED IMAGE INTENSIFIER FOR X-RAY-DIFFRACTION
    AMEMIYA, Y
    ITO, K
    YAGI, N
    ASANO, Y
    WAKABAYASHI, K
    UEKI, T
    ENDO, T
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02) : 2290 - 2294
  • [2] AMENISCH H, 1999, 11 INT C SMALL ANGL, P239
  • [3] SYSTEMS FOR TIME RESOLVED X-RAY MEASUREMENTS USING ONE-DIMENSIONAL AND TWO-DIMENSIONAL DETECTORS - REQUIREMENTS AND PRACTICAL EXPERIENCE
    BOULIN, C
    DAINTON, D
    DORRINGTON, E
    ELSNER, G
    GABRIEL, A
    BORDAS, J
    KOCH, MHJ
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 201 (01): : 209 - 220
  • [4] BRAS W, 1999, 11 INT C SMALL ANGL, P272
  • [5] FOWLER AG, 1983, J BIOCHEM BIOPH METH, V7, P387
  • [6] FUJISAWA T, 1999, 11 INT C SMALL ANGL, P276
  • [7] FUJISAWA T, 1997, SPRING 8 ANN REPORT, P238
  • [8] POSITION-SENSITIVE X-RAY DETECTOR
    GABRIEL, A
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1977, 48 (10) : 1303 - 1305
  • [9] CALIBRATION AND CORRECTION OF SPATIAL DISTORTIONS IN 2D DETECTOR SYSTEMS
    HAMMERSLEY, AP
    SVENSSON, SO
    THOMPSON, A
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 346 (1-2) : 312 - 321
  • [10] Comparison of the three-dimensional structures of recombinant human H and horse L ferritins at high resolution
    Hempstead, PD
    Yewdall, SJ
    Fernie, AR
    Lawson, DM
    Artymiuk, PJ
    Rice, DW
    Ford, GC
    Harrison, PM
    [J]. JOURNAL OF MOLECULAR BIOLOGY, 1997, 268 (02) : 424 - 448