Biologic width around titanium implants. A histometric analysis of the implanto-gingival junction around unloaded and loaded nonsubmerged implants in the canine mandible

被引:403
作者
Cochran, DL
Hermann, JS
Schenk, RK
Higginbottom, FL
Buser, D
机构
[1] UNIV BERN, INST PATHOPHYSIOL, CH-3012 BERN, SWITZERLAND
[2] BAYLOR COLL DENT, DEPT ORAL & MAXILLOFACIAL SURG & PHARMACOL, DALLAS, TX 75246 USA
[3] UNIV BERN, SCH DENT MED, DEPT ORAL SURG, CH-3012 BERN, SWITZERLAND
关键词
dental implants endosseous; biologic width; titanium; soft tissue; animal studies;
D O I
10.1902/jop.1997.68.2.186
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
THE USE OF ENDOSSEOUS DENTAL IMPLANTS as transmucosal devices necessitates the successful integration of three different tissues: bone, connective tissue, and epithelium. So far, studies have predominantly focused on hard tissue integration. Much less is known about soft tissues. This study examined the dimensions of the implanto-gingival-junction in relation to clinically healthy unloaded and loaded nonsubmerged implants. In total, 69 titanium plasma-sprayed (TPS) and sandblasted acid-etched (SLA) implants were placed in an alternating fashion in six foxhounds and allowed to heal for 3 months. Two dogs were sacrificed after the initial healing period. The remaining four dogs had crowns fabricated that were allowed to function for up to 12 months. These animals were sacrificed after 3 and 12 months of loading. Histometric analysis of undecalcified histologic sections included the evaluation of the sulcus depth (SD), the dimensions of the junctional epithelium (JE), and the connective tissue contact (CTC). Mean values in the 3 month unloaded group were 0.49 mm for SD, 1.16 mm for JE, and 1.36 mm for CTC. These dimensions were 0.50 mm for SD, 1.44 mm for JE, and 1.01 mm for CTC for the 3 month loaded group. After 12 months of loading, these values were 0.16 mm for SD, 1.88 mm for JE, and 1.05 mm for CTC. The sum of these measurements was similar for the different time points and similar to the same dimensions around teeth. TPS and SLA surfaces had no influence on the evaluated parameters (P > 0.05). The data suggest that a biologic width exists around unloaded and loaded nonsubmerged one-part titanium implants and that this is a physiologically formed and stable dimension as is found around teeth.
引用
收藏
页码:186 / 197
页数:12
相关论文
共 69 条
  • [1] MARGINAL TISSUE-REACTIONS AT OSSEOINTEGRATED TITANIUM FIXTURES .1. A 3-YEAR LONGITUDINAL PROSPECTIVE-STUDY
    ADELL, R
    LEKHOLM, U
    ROCKLER, B
    BRANEMARK, PI
    LINDHE, J
    ERIKSSON, B
    SBORDONE, L
    [J]. INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1986, 15 (01) : 39 - 52
  • [2] A 15-YEAR STUDY OF OSSEOINTEGRATED IMPLANTS IN THE TREATMENT OF THE EDENTULOUS JAW
    ADELL, R
    LEKHOLM, U
    ROCKLER, B
    BRANEMARK, PI
    [J]. INTERNATIONAL JOURNAL OF ORAL SURGERY, 1981, 10 (06): : 387 - 416
  • [3] ADELL R, 1990, International Journal of Oral and Maxillofacial Implants, V5, P347
  • [4] AHLQVIST J, 1990, International Journal of Oral and Maxillofacial Implants, V5, P155
  • [5] Albrektsson T, 1986, Int J Oral Maxillofac Implants, V1, P11
  • [6] STATE-OF-THE-ART IN ORAL IMPLANTS
    ALBREKTSSON, T
    SENNERBY, L
    [J]. JOURNAL OF CLINICAL PERIODONTOLOGY, 1991, 18 (06) : 474 - 481
  • [7] DIRECT BONE ANCHORAGE OF DENTAL IMPLANTS
    ALBREKTSSON, T
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1983, 50 (02) : 255 - 261
  • [8] TITANIUM PLASMA-SPRAYED (TPS) SCREW IMPLANTS FOR THE RECONSTRUCTION OF THE EDENTULOUS MANDIBLE
    BABBUSH, CA
    KENT, JN
    MISIEK, DJ
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1986, 44 (04) : 274 - 282
  • [9] BEHNEKE A, 1992, Z ZAHNARZTLICHE IMPL, V8, P97
  • [10] The soft tissue barrier at implants and teeth
    Berglundh, T.
    Lindhe, J.
    Ericsson, I.
    Marinello, C. P.
    Liljenberg, B.
    Thomsen, P.
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 1991, 2 (02) : 81 - 90