Improved image quality and detection of acute cerebral infarction with PROPELLER diffusion-weighted MR imaging

被引:85
作者
Forbes, KP [1 ]
Pipe, JG [1 ]
Karis, JP [1 ]
Heiserman, JE [1 ]
机构
[1] St Josephs Hosp, Barrow Neurol Inst, Div Neuroradiol, Phoenix, AZ 85013 USA
关键词
brain; infarction; magnetic resonance (MR); diffusion tudy; k-space; pulse sequences;
D O I
10.1148/radiol.2252011479
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To compare periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) multishot fast spin-echo diffusion-weighted magnetic resonance (MR) imaging with single-shot echo-planar diffusion-weighted MR imaging for image quality and visualization of acute cerebral infarction. MATERIALS AND METHODS: Seventy subjects (35 men, 35 women; mean age, 55 years 24 [SD]) who were suspected of having acute cerebral infarction (symptom duration, 2.8 days :2.7) underwent PROPELLER and echo-planar MR imaging (b = 1,000 sec/mm(2)). Two neuroradiologists compared unlabeled images for presence of artifacts, visualization of infarction, and their preference of images. Interobserver agreement and image comparison were assessed by using the K statistic and the x(2) test, respectively. RESULTS: PROPELLER MR imaging reduced susceptibility artifacts (n = 70 subjects), limited visualization of temporal (echo-planar, n = 64; PROPELLER, n = 0; P < which 2.01, X test), frontal (echo-planar, n = 58; PROPELLER, n = 1; P <.01), and parietal n = 5; PROPELLER, n = 0; P <.05) and cerebellum (echo-planar, lobes (echo-planar, n = 0; P <.01) and brainstem (echo-planar, n = 23; n = 36; PROPELLER, PROPELLER, n = 0;,P <.01). Acute infarction (n = 31 subjects) was better demonstrated at PROPELLER MR imaging (PROPELLER better, n = 18; echo-planar better, n = 1; PROPELLER and echo-planar equal, n = 12; P <.01, x(2) test). PROPELLER MR imaging was preferred in all (n = 70) but one case in which the lesion lay within the intersection gap (PROPELLER preferred, n = 69; echo-planar preferred, n = 1; P <.01, x(2) test). CONCLUSION: With a short increase in imaging time, PROPELLER MR imaging offers better image quality and detection of acute cerebral infarction than does echo-planar MR imaging. (C) RSNA, 2002.
引用
收藏
页码:551 / 555
页数:5
相关论文
共 10 条
[1]   PROPELLER MRI: Clinical testing of a novel technique for quantification and compensation of head motion [J].
Forbes, KPN ;
Pipe, JG ;
Bird, CR ;
Heiserman, JE .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (03) :215-222
[2]   DENTAL MATERIALS AND MAGNETIC-RESONANCE-IMAGING [J].
LISSAC, M ;
METROP, D ;
BRUGIRARD, J ;
COUDERT, JL ;
PIMMEL, P ;
BRIGUET, A ;
REVEL, D ;
AMIEL, M .
INVESTIGATIVE RADIOLOGY, 1991, 26 (01) :40-45
[3]  
MOSELEY ME, 1990, AM J NEURORADIOL, V11, P423
[4]   Anatomic localization and quantitative analysis of gradient refocused echo-planar fMRI susceptibility artifacts [J].
Ojemann, JG ;
Akbudak, E ;
Snyder, AZ ;
McKinstry, RC ;
Raichle, ME ;
Conturo, TE .
NEUROIMAGE, 1997, 6 (03) :156-167
[5]  
Pipe JG, 1999, MAGNET RESON MED, V42, P963, DOI 10.1002/(SICI)1522-2594(199911)42:5<963::AID-MRM17>3.0.CO
[6]  
2-L
[7]   Multishot diffusion-weighted FSE using PROPELLER MRI [J].
Pipe, JG ;
Farthing, VG ;
Forbes, KP .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (01) :42-52
[8]   Aneurysm clips: Evaluation of MR imaging artifacts at 1.5 T [J].
Shellock, FG ;
Kanal, E .
RADIOLOGY, 1998, 209 (02) :563-566
[9]   ECHO-PLANAR IMAGING OF INTRAVOXEL INCOHERENT MOTION [J].
TURNER, R ;
LEBIHAN, D ;
MAIER, J ;
VAVREK, R ;
HEDGES, LK ;
PEKAR, J .
RADIOLOGY, 1990, 177 (02) :407-414
[10]   Diffusion-weighted magnetic resonance imaging in acute stroke [J].
van Everdingen, KJ ;
van der Grond, J ;
Kappelle, LJ ;
Ramos, LMP ;
Mali, WPTM .
STROKE, 1998, 29 (09) :1783-1790