Serum thyroxine binding capacity-dependent bias in an automated free thyroxine assay

被引:18
作者
Christofides, ND
Wilkinson, E
Stoddart, M
Ray, DC
Beckett, GJ
机构
[1] Cardiff Labs, Res & Dev, Cardiff CF4 7YT, S Glam, Wales
[2] Univ Edinburgh, Royal Infirm, Dept Clin Biochem, Edinburgh EH3 9YW, Midlothian, Scotland
来源
JOURNAL OF IMMUNOASSAY | 1999年 / 20卷 / 04期
关键词
AXSYM free thyroxine; binding capacity; equilibrium dialysis; protein-bound T4; immunoassay;
D O I
10.1080/01971529909349351
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
The magnitude of serum thyroxine (T4) binding capacity (sBC) dependent bias in the AXSYM free thyroxine (FT4) assay was assessed using two recently described tests. One of the tests uses a direct equilibrium dialysis (ED) FT4 assay as the reference method. The results obtained with the AXSYM method were compared with those obtained by the ED FT4 method in patient sera having a wide range of sBC The other test involves comparison of the FT4 results obtained following dilution of sera by an inert buffer, to theoretically derived FT4 results. As serum dilution causes a predictable decrease in sBC, the demonstration of a negative bias whose magnitude increases in parallel to the dilution, is indicative of an sBC-dependent bias. The AXSYM FT4 assay exhibited a significant sBC-dependent bias. This sBC-dependent bias is likely to have been caused by the presence of significant amounts of T4 binding proteins in the assay reagents.