Improved estimation of the limit of detection in multivariate calibration

被引:16
作者
Faber, K [1 ]
Kowalski, BR [1 ]
机构
[1] UNIV WASHINGTON,CTR PROC ANALYT CHEM,SEATTLE,WA 98195
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1997年 / 357卷 / 07期
关键词
D O I
10.1007/s002160050250
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The limit of detection is one of the most important performance characteristics of an analytical procedure. This paper critically examines a recently introduced expression for estimating a multivariate limit of detection. The main result is that the proposed expression is inconsistent. An application from inductively coupled plasma - optical emission spectrometry (ICP-OES) shows that the true limit of detection may be overestimated by more than 20%. In addition, the practical evaluation of the proposed expression amounts to an iterative procedure, which is unattractive from an application point of view. A modification is derived that solves both problems. The relevant expressions are discussed with respect to their interpretability in terms of experimental design.
引用
收藏
页码:789 / 795
页数:7
相关论文