Planar gradient coil design by scaling the spatial frequencies of minimum-inductance current density

被引:15
作者
Lee, SY [1 ]
Park, BS [1 ]
Yi, JH [1 ]
Yi, W [1 ]
机构
[1] SAMSUNG ADV INST TECHNOL,MED EQUIPMENT LAB,CHOONGBUK,SOUTH KOREA
关键词
planar gradient coil; magnetic energy minimization; spatial frequency scaling; linearity error;
D O I
10.1002/mrm.1910380525
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Gradient coil inductance has been remarkably reduced by the minimum-inductance design technique, which minimizes the magnetic energy stored by the gradient coil. The planar gradient coil designed by this technique, however, often has poor magnetic field linearity. Scaling the spatial frequencies of the current density function derived by this method, the magnetic field linearity of the planar gradient coil can be greatly improved with a small sacrifice of gradient coil inductance, A figure of merit of the planar gradient coil has been found to be improved by scaling the spatial frequencies.
引用
收藏
页码:858 / 861
页数:4
相关论文
共 10 条
[1]   MAGNETIC-FIELD GRADIENT COILS FOR NMR IMAGING [J].
BANGERT, V ;
MANSFIELD, P .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1982, 15 (02) :235-239
[2]   SCREENED COIL DESIGNS FOR NMR IMAGING IN MAGNETS WITH TRANSVERSE FIELD GEOMETRY [J].
BOWTELL, R ;
MANSFIELD, P .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1990, 1 (05) :431-439
[3]   ACTIVE MAGNETIC SCREENING OF COILS FOR STATIC AND TIME-DEPENDENT MAGNETIC-FIELD GENERATION IN NMR IMAGING [J].
MANSFIELD, P ;
CHAPMAN, B .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1986, 19 (07) :540-545
[4]   INSERTABLE BIPLANAR GRADIENT COILS FOR MAGNETIC-RESONANCE-IMAGING [J].
MARTENS, MA ;
PETROPOULOS, LS ;
BROWN, RW ;
ANDREWS, JH ;
MORICH, MA ;
PATRICK, JL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (11) :2639-2645
[5]  
PETROPOULOS LS, 1995, P SMR 3 ANN M NIC, P942
[6]   IMPROVING MAGNETIC-FIELD GRADIENT COILS FOR NMR IMAGING [J].
SUITS, BH ;
WILKEN, DE .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1989, 22 (08) :565-573
[7]   MINIMUM INDUCTANCE COILS [J].
TURNER, R .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1988, 21 (10) :948-952
[8]   GRADIENT COIL DESIGN - A REVIEW OF METHODS [J].
TURNER, R .
MAGNETIC RESONANCE IMAGING, 1993, 11 (07) :903-920
[9]   COIL OPTIMIZATION FOR MRI BY CONJUGATE-GRADIENT DESCENT [J].
WONG, EC ;
JESMANOWICZ, A ;
HYDE, JS .
MAGNETIC RESONANCE IN MEDICINE, 1991, 21 (01) :39-48
[10]   ANALYTICAL DESIGN METHOD OF SELF-SHIELDED PLANAR COILS [J].
YODA, K .
JOURNAL OF APPLIED PHYSICS, 1990, 67 (09) :4349-4353