Ethamsylate (Dicynone®) Interference in Determination of Serum Creatinine, Uric Acid, Triglycerides, and Cholesterol in Assays Involving the Trinder Reaction; In Vivo and In Vitro

被引:15
作者
Dastych, Milan [1 ,2 ]
Wiewiorka, Ondkej [1 ]
Benovska, Miroslava [1 ,2 ]
机构
[1] Univ Hosp Brno, Dept Clin Biochem, Brno 62500, Czech Republic
[2] Masaryk Univ, Fac Med, Dept Lab Methods, Brno, Czech Republic
关键词
trinder reaction; interference; ethamsylate; PEROXIDASE;
D O I
10.7754/Clin.Lab.2013.130902
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
100118 [医学信息学]; 100208 [临床检验诊断学];
摘要
Background: The aim of our research was the quantification of interfering properties of the haemostatic drug Dicynone (R) (ethamsylate) in serum creatinine, uric acid, cholesterol, and triglyceride assays using the Trinder reaction. Methods: Blood from patients was collected before and 15 minutes after administration of 500 mg Dicynone (R) dose i.v. and the above mentioned analytes were quantified using Roche assays (Cobas 8000). In our in vitro experiment, we measured concentrations of the analytes in pooled serum aliquots with final concentrations of Dicynone (R) additions 0, 30, 60, 150, and 300 mg/L. Aliquots with 60 mg/L Dicynone (R) were also measured at 2, 6, and 8 hours after initial measurement when stored in 22 degrees C and 4 degrees C for comparison. Results: Concentrations of the measured analytes in samples from patients administered with a 500 mg dose of Dicynone (R) were lower in all cases (n = 10) when compared to values in samples taken immediately before treatment. The in vitro samples showed that considerable negative interference occurred even with the low concentrations of Dicynone (R) additions (30 and 60 mg/L), showing the strongest negative interference in creatinine values, followed by uric acid, triglycerides, and cholesterol. Using in vitro samples, we showed strong time and temperature dependence on Dicynone (R) interference. Conclusions: We found and proved significant negative interference of the drug Dicynone (R) (ethamsylate) in the clinical analysis of blood using in vivo and in vitro experiments. Furthermore, we observed a change of this effect in serum matrix over time and at different storage temperatures.
引用
收藏
页码:1373 / 1376
页数:4
相关论文
共 6 条
[1]
Chromogenic System for Measuring Hydrogen Peroxide: The Enzymatic Uric Acid Assay [J].
Fossati, Piero ;
Prencipe, Lorenzo .
CLINICAL CHEMISTRY, 2010, 56 (05) :865-866
[2]
FUSHIMI R, 1992, CLIN CHEM, V38, P169
[3]
HCP-MEPPO, DIC 250 INJ SOL OM P
[4]
Jiang Y-F, 2008, CHIN J LAB DIAGN, V3, P37
[5]
Karon BS, 1998, CLIN CHEM, V44, P155
[6]
SPAIN MA, 1986, CLIN CHEM, V32, P518