Measurement of the x-ray mass energy-absorption coefficient of air using 3 keV to 10 keV synchrotron radiation

被引:31
作者
Bueermann, L.
Grosswendt, B.
Kramer, H-M
Selbach, H-J
Gerlach, M.
Hoffmann, M.
Krumrey, M.
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
[2] Phys Tech Bundesanstalt, D-10587 Berlin, Germany
关键词
D O I
10.1088/0031-9155/51/20/004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For the first time absolute photon mass energy-absorption coefficients of air in the energy range 3 keV to 10 keV have been measured with relative standard uncertainties less than 1%, significantly smaller than those of up to 5% assumed hitherto for calculated data. Monochromatized synchrotron radiation was used to measure both the total radiant energy by means of silicon photodiodes calibrated against a cryogenic radiometer and the fraction of radiant energy that is deposited in dry air by means of a free air ionization chamber. The measured ionization charge was converted into energy absorbed in air by calculated effective W values of photons as a function of their energy based on new measurements of the W values in dry air for electron kinetic energies between 1 keV and 7 keV, also presented in this work. The measured absorption coefficients were compared with state-of-the art calculations and found to agree within 0.7% with data calculated earlier by Hubbell at energies above 4 keV but were found to differ by values up to 2.1% at 10 keV from more recent calculations of Seltzer.
引用
收藏
页码:5125 / 5150
页数:26
相关论文
共 25 条
[1]  
[Anonymous], 1987, RAD DOSIMETRY
[2]  
[Anonymous], 1995, GUID EXPR UNC MEAS
[3]   PARTITION OF KERMA TO ACCOUNT FOR BREMSSTRAHLUNG [J].
ATTIX, FH .
HEALTH PHYSICS, 1979, 36 (03) :347-354
[4]   Brachytherapy dosimetry of 125I and 103Pd sources using an updated cross section library for the MCNP Monte Carlo transport code [J].
Bohm, TD ;
DeLuca, PM ;
DeWerd, LA .
MEDICAL PHYSICS, 2003, 30 (04) :701-711
[5]   ATOMIC INNER-SHELL LEVEL ENERGIES DETERMINED BY ABSORPTION SPECTROMETRY WITH SYNCHROTRON RADIATION [J].
BREINIG, M ;
CHEN, MH ;
ICE, GE ;
PARENTE, F ;
CRASEMANN, B .
PHYSICAL REVIEW A, 1980, 22 (02) :520-528
[6]  
BUEMANN L, 2003, INT S STAND COD PRAC, V1, P125
[7]  
CULLEN DE, 1997, UCRL50400 L LIV NAT, V6
[8]   An analysis of MCNP cross-sections and tally methods for low-energy photon emitters [J].
DeMarco, JJ ;
Wallace, RE ;
Boedeker, K .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (08) :1321-1332
[9]  
ENGELKE BA, 1988, PTBDOS16
[10]  
EVANS RD, 1958, HDB PHYSIK, V34, P218