An improved neutron collimator for brain tumor irradiations in clinical boron neutron capture therapy

被引:52
作者
Liu, HB [1 ]
Greenberg, DD [1 ]
Capala, J [1 ]
Wheeler, FJ [1 ]
机构
[1] IDAHO NATL ENGN LAB, IDAHO FALLS, ID 83415 USA
关键词
BNCT; epithermal neutron beam; collimator; Monte Carlo; dosimetry;
D O I
10.1118/1.597772
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To improve beam penetration into a head allowing the treatment of deeper seated tumors, two neutron collimators were built sequentially and tested for use in the clinical boron neutron capture therapy (BNCT) program at the epithermal neutron irradiation facility of the Brookhaven Medical Research Reactor. The collimators were constructed from lithium-impregnated polyethylene, which comprises Li2CO3 powder (approximate to 93% enriched isotopic Li-6) uniformly dispersed in polyethylene to a total Li-6 content of 7.0 wt.%. The first collimator is 7.6 cm thick with a conical cavity 16 cm in diameter on the reactor core side tapering to 8 cm facing the patient's head. The second collimator is 15.2 cm thick with a conical cavity 20 cm in diameter tapering to 12 cm. A clinical trial of BNCT for patients with malignant brain tumors is underway using the first collimator. Results of phantom dosimetry and Monte Carlo computations indicate that the new 15.2 cm thick collimator will improve the neutron beam penetration. Thus, the second collimator was made and will be used in an upcoming clinical trial. In-air and in-phantom mixed-field dosimetric measurements were compared to Monte Carlo computations for both collimators. The deeper penetration is achieved but at a sacrifice in beam intensity. In this report, a performance comparison of both collimators regarding various fluence rate and absorbed dose distributions in a head model is presented and discussed. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:2051 / 2060
页数:10
相关论文
共 24 条
[1]  
ATTIX F. H., 1986, Introduction to radiological physics and radiation dosimetry
[2]  
BARTH RF, 1990, CANCER RES, V50, P1061
[3]  
BRIESMEISTER JF, 1986, 7396M LOS AL NAT LAB
[4]   EXPLANATION OF CEREBRAL WHITE-GRAY CONTRAST IN COMPUTED-TOMOGRAPHY [J].
BROOKS, RA ;
DICHIRO, G ;
KELLER, MR .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1980, 4 (04) :489-491
[5]  
CHANANA AD, 1994, BORON NEUTRON CAPTUR
[6]  
CHANANA AD, 1996, BORON NEUTRON CAPTUR
[7]   NEUTRON-CAPTURE THERAPY OF THE 9L RAT GLIOSARCOMA USING THE P-BORONOPHENYLALANINE-FRUCTOSE COMPLEX [J].
CODERRE, JA ;
BUTTON, TM ;
MICCA, PL ;
FISHER, CD ;
NAWROCKY, MM ;
LIU, HB .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 30 (03) :643-652
[8]   DERIVATIONS OF RELATIVE BIOLOGICAL EFFECTIVENESS FOR THE HIGH-LET RADIATIONS PRODUCED DURING BORON NEUTRON-CAPTURE IRRADIATIONS OF THE 9L RAT GLIOSARCOMA IN-VITRO AND IN-VIVO [J].
CODERRE, JA ;
MAKAR, MS ;
MICCA, PL ;
NAWROCKY, MM ;
LIU, HYB ;
JOEL, DD ;
SLATKIN, DN ;
AMOLS, HI .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (05) :1121-1129
[9]  
CODERRE JA, IN PRESS J NEUROONCO
[10]  
FAIRCHILD RG, 1990, BASIC LIFE SCI, V54, P185