Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla

被引:299
作者
Martin, ET
Coman, JA
Shellock, FG
Pulling, CC
Fair, R
Jenkins, K
机构
[1] Oklahoma Heart Inst, Tulsa, OK 74133 USA
[2] Univ Oklahoma, Tulsa, OK USA
[3] Integra CTS, Minneapolis, MN USA
[4] Inst Magnet Resonance Safety Educ & Res, Los Angeles, CA USA
[5] Univ So Calif, Keck Sch Med, Los Angeles, CA USA
关键词
D O I
10.1016/j.jacc.2003.12.016
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES The study was done to determine whether patients with pacemakers could safely undergo magnetic resonance imaging (MRI) at 1.5-Tesla (T). BACKGROUND Because of theoretical risks, it is an absolute contraindication for a patient with a pacemaker to undergo MRI. However, there are times when an MRI is needed to provide valuable clinical information. METHODS Fifty-four patients underwent a total of 62 MRI examinations at 1.5-T. The type of MRI examination was not limited and included cardiac, vascular, and general MRI studies using various whole-body averaged specific absorption rate (SAR) of radiofrequency power. Restrictions were not placed on the type of pacemaker present in the patient. All pacemakers were interrogated immediately before and after MRI scanning, and patients were continuously monitored. Before and after MRI, interrogation was done, and pacing and sensing thresholds, as well as lead impedances, were all measured. RESULTS A total of 107 leads and 61 pulse generators were evaluated. No adverse events occurred. Forty (37%) of the leads underwent changes, whereas 10 (9.4%) leads underwent a significant change. Only 2 of the 107 (1.9%) leads required a change in programmed output. Threshold changes were unrelated to cardiac chamber, anatomical location, peak SAR, and time from lead implant to the MRI examination. Electrocardiographic changes and patient symptoms were minor and did not require cessation of MRI. CONCLUSIONS Safety was demonstrated in this series of patients with pacemakers at 1.5-T. (C) 2004 by the American College of Cardiology Foundation.
引用
收藏
页码:1315 / 1324
页数:10
相关论文
共 48 条
[1]   Effects of magnetic resonance Imaging on cardiac pacemakers and electrodes [J].
Achenbach, S ;
Moshage, W ;
Diem, B ;
Bieberle, T ;
Schibgilla, V ;
Bachmann, K .
AMERICAN HEART JOURNAL, 1997, 134 (03) :467-473
[2]   NUCLEAR MAGNETIC-RESONANCE IMAGING IN A PATIENT WITH A DDD PACEMAKER [J].
ALAGONA, P ;
TOOLE, JC ;
MANISCALCO, BS ;
GLOVER, MU ;
ABERNATHY, GT ;
PRIDA, XE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (04) :619-619
[3]  
Avery J K, 1988, J Tenn Med Assoc, V81, P523
[4]   ABSENCE OF RADIOFREQUENCY HEATING FROM AUDITORY IMPLANTS DURING MAGNETIC-RESONANCE-IMAGING [J].
CHOU, CK ;
MCDOUGALL, JA ;
CHAN, KW .
BIOELECTROMAGNETICS, 1995, 16 (05) :307-316
[5]  
CHOU CK, 1997, IEEE T BIOMED ENG, V44, P357
[6]  
Coman JA, 2001, J AM COLL CARDIOL, V37, p436A
[7]   Pacing in magnetic resonance imaging environment:: Clinical and technical considerations on compatibility [J].
Duru, F ;
Luechinger, R ;
Scheidegger, MB ;
Lüscher, TF ;
Boesiger, P ;
Candinas, R .
EUROPEAN HEART JOURNAL, 2001, 22 (02) :113-124
[8]   EFFECT OF MAGNETIC-RESONANCE-IMAGING ON DDD PACEMAKERS [J].
ERLEBACHER, JA ;
CAHILL, PT ;
PANNIZZO, F ;
KNOWLES, RJR .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (06) :437-440
[9]   THE EFFECTS OF NUCLEAR MAGNETIC-RESONANCE IMAGERS ON EXTERNAL AND IMPLANTABLE PULSE GENERATORS [J].
FETTER, J ;
ARAM, G ;
HOLMES, DR ;
GRAY, JE ;
HAYES, DL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (04) :720-727
[10]  
Finelli DA, 2002, AM J NEURORADIOL, V23, P1795