Thermal conductivity of reference solid materials

被引:43
作者
Assael, MJ [1 ]
Gialou, K [1 ]
Kakosimos, K [1 ]
Metaxa, I [1 ]
机构
[1] Aristotle Univ Thessaloniki, Dept Chem Engn, Thessaloniki 54124, Greece
关键词
AISI; 304L; Pyrex; 7740; Pyroceram; 9606; thermal conductivity; transient hot wire;
D O I
10.1023/B:IJOT.0000028477.74595.d5
中图分类号
O414.1 [热力学];
学科分类号
摘要
The thermal conductivity of three thermal-conductivity reference materials, Pyrex 7740, Pyroceram 9606, and stainless steel AISI 304L, has been studied. The technique employed is the transient hot-wire technique, and measurements cover a temperature range from room temperature up to 570 K. The technique is applied here in a novel way that eliminates all remaining contact resistances. This allows the apparatus to operate in an absolute way. The method makes use of a soft silicone paste material between the hot wires of the technique and the solid of interest. Measurements of the transient temperature rise of the wires in response to an electrical heating step in the wires over a period of 20 mus up to 20 s allow an absolute determination of the thermal conductivity of the solid, as well as of the silicone paste. The method is based on a full theoretical model with equations solved by a two-dimensional finite-element method applied to the exact geometry. At the 95% confidence level, the standard deviation of the thermal conductivity measurements is 0.1% for Pyrex 7740, 0.4% for Pyroceram 9606, and 0.2% for stainless steel AISI 304L, while the standard uncertainty of the technique is less than 1.5%.
引用
收藏
页码:397 / 408
页数:12
相关论文
共 16 条
[1]  
[Anonymous], 1995, GUID EXPR UNC MEAS
[2]   A transient hot-wire instrument for the measurement of the thermal conductivity of solids up to 590 K [J].
Assael, MJ ;
Gialou, K .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2003, 24 (03) :667-674
[3]   Measurement of the thermal conductivity of stainless steel AISI 304L up to 550 K [J].
Assael, MJ ;
Gialou, K .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2003, 24 (04) :1145-1153
[4]   Application of the transient hot-wire technique to the measurement of the thermal conductivity of solids [J].
Assael, MJ ;
Dix, M ;
Gialou, K ;
Vozar, L ;
Wakeham, WA .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2002, 23 (03) :615-633
[5]  
Bogaard R.H., 1985, Thermal Conductivity, V18, P175
[6]  
Chu T.K., 1977, THERMAL CONDUCTIVITY, V15, P79
[7]  
*CRM, 1990, 039 CRM BCR COMM EUR
[8]   THE THERMAL-CONDUCTIVITY OF AISI 304L STAINLESS-STEEL [J].
GRAVES, RS ;
KOLLIE, TG ;
MCELROY, DL ;
GILCHRIST, KE .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 1991, 12 (02) :409-415
[9]   TRANSIENT PLANE SOURCE TECHNIQUES FOR THERMAL-CONDUCTIVITY AND THERMAL-DIFFUSIVITY MEASUREMENTS OF SOLID MATERIALS [J].
GUSTAFSSON, SE .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (03) :797-804
[10]  
HULSTROM LC, 1988, THERM CONDUCT, V19, P199