Thermal calorimetric detector for atomic hydrogen and deuterium beam measurements

被引:5
作者
Ershov, VP [1 ]
Fimushkin, VV [1 ]
Kulikov, MV [1 ]
Pilipenko, YK [1 ]
Shutov, VB [1 ]
机构
[1] JOINT INST NUCL RES,RU-141980 DUBNA,MOSCOW REGION,RUSSIA
关键词
D O I
10.1016/S0168-9002(96)00759-0
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A method for atomic hydrogen and deuterium beam intensity measurements by means of a low temperature thermal calorimetric detector due to beam atom recombination is discussed. It is shown that the method could have some advantages, in particular experimental conditions with respect to a conventional compression tube method especially for pulsed beams. The efficiency of recombination energy accommodation for atomic hydrogen and deuterium beams is experimentally determined. The effective efficiency is about 11% for hydrogen and about 18% for deuterium. Some applications of the method are presented.
引用
收藏
页码:435 / 437
页数:3
相关论文
共 10 条
[1]   CONDENSATION OF ATOMIC AND MOLECULAR HYDROGEN AT LOW TEMPERATURES [J].
BRACKMANN, R ;
FITE, WL .
JOURNAL OF CHEMICAL PHYSICS, 1961, 34 (05) :1572-&
[2]  
DATSKOV VI, 1684677 JINR
[3]  
ERHSOV VP, 1994, P 8 INT S POL PHEN N, P610
[4]  
ERSHOV VP, E1390331 JINR
[5]  
ERSHOV VP, 1993, P 5 WORKSH HIGH EN S, P384
[6]   PHYSICAL MOTIVATIONS FOR THERMAL DETECTORS [J].
FIORINI, E .
JOURNAL OF LOW TEMPERATURE PHYSICS, 1993, 93 (3-4) :189-200
[7]  
KETZEL M, E892155 JINR
[8]  
MARNECO G, 1972, J VAC SCI TECHNOL, V9, P824
[9]  
NEGANOV B, 1978, JETP LETT, V28, P258
[10]   RECOMBINATION OF ATOMIC-HYDROGEN ON LOW-TEMPERATURE SURFACES [J].
SCHUTTE, A ;
BASSI, D ;
TOMMASINI, F ;
TURELLI, A ;
SCOLES, G ;
HERMANS, LJF .
JOURNAL OF CHEMICAL PHYSICS, 1976, 64 (10) :4135-4142