Development of a haemocompatible pO(2) sensor with phospholipid-based copolymer membrane

被引:29
作者
Zhang, SF
Benmakroha, Y
Rolfe, P
Tanaka, S
Ishihara, K
机构
[1] YAMAGATA UNIV,FAC ENGN,DEPT ELECT & INFORMAT ENGN,YONEZAWA,YAMAGATA 992,JAPAN
[2] TOKYO MED & DENT UNIV,INST MED & DENT ENGN,TOKYO 101,JAPAN
基金
英国工程与自然科学研究理事会;
关键词
intravascular sensor; polymer membrane; phospholipid; haemocompatibility; in vitro test; blood contamination; long-term performance;
D O I
10.1016/0956-5663(96)87661-5
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This paper describes research to improve the haemocompatibility of an intravascular pO(2) sensor utilizing a novel copolymer membrane. This material is based on a methacrylate monomer having an electrically neutral head group, 2-methacryloyloxyethyl phosphorylcholine (MPC), copolymerized with n-butyl methacrylate (BMA). The MPC mimics the outer layer lipid structure of the red blood cell membrane, and thus poly (MPC-co-BMA) could be expected to show excellent compatibility with blood. In this paper the methods of preparation of a pO(2) sensor covered with different types of membrane, including polyetherurethane (PU), polyvinyl chloride (PVC), poly (MPC-co-BMA) and a compound membrane [PVC+poly (MPC-co-BMA)] are described. The surface chemical structure of the different types of membrane were characterized with Fourier transform infrared spectroscopy. Scanning electron microscopy was used to examine the surface contamination after the sensors had been used in blood. The properties of the sensors were first assessed in phosphate-buffered solution, and the short-term and long-term performances were then demonstrated by use of buffer with mixed proteins, serum and whole human blood. The sensors coated with poly (MPC-co-BMA) and PVC+poly (MPC-co-BMA) membranes showed stable behaviour in static human blood, whereas the sensors with PU and PVC membrane were subject to a continuous signal deterioration during the test period. (C) 1996 Elsevier Science Limited.
引用
收藏
页码:1019 / 1029
页数:11
相关论文
共 24 条
[1]   A BIOCOMPATIBLE NEEDLE-TYPE GLUCOSE SENSOR BASED ON PLATINUM-ELECTROPLATED CARBON ELECTRODE [J].
CHEN, CY ;
TAMIYA, E ;
ISHIHARA, K ;
KOSUGI, Y ;
SU, YC ;
NAKABAYASHI, N ;
KARUBE, I .
APPLIED BIOCHEMISTRY AND BIOTECHNOLOGY, 1992, 36 (03) :211-226
[2]   ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY [J].
CLARK, LC ;
LYONS, C .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1962, 102 (01) :29-&
[3]  
DURRANI AA, 1987, POLYM SURFACES INTER, P189
[4]   ROTATED, MEMBRANE-COVERED OXYGEN-ELECTRODE [J].
GOUGH, DA ;
LEYPOLDT, JK .
ANALYTICAL CHEMISTRY, 1980, 52 (07) :1126-1130
[5]   BIOMEMBRANES AS MODELS FOR POLYMER SURFACES .5. THROMBOELASTOGRAPHIC STUDIES OF POLYMERIC LIPIDS AND POLYESTERS [J].
HALL, B ;
BIRD, RL ;
KOJIMA, M ;
CHAPMAN, D .
BIOMATERIALS, 1989, 10 (04) :219-224
[6]   BIOMEMBRANE SURFACES AS MODELS FOR POLYMER DESIGN - THE POTENTIAL FOR HEMOCOMPATIBILITY [J].
HAYWARD, JA ;
CHAPMAN, D .
BIOMATERIALS, 1984, 5 (03) :135-142
[7]   REDUCED THROMBOGENICITY OF POLYMERS HAVING PHOSPHOLIPID POLAR GROUPS [J].
ISHIHARA, K ;
ARAGAKI, R ;
UEDA, T ;
WATENABE, A ;
NAKABAYASHI, N .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1990, 24 (08) :1069-1077
[8]   HEMOCOMPATIBILITY OF HUMAN WHOLE-BLOOD ON POLYMERS WITH A PHOSPHOLIPID POLAR GROUP AND ITS MECHANISM [J].
ISHIHARA, K ;
OSHIDA, H ;
ENDO, Y ;
UEDA, T ;
WATANABE, A ;
NAKABAYASHI, N .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (12) :1543-1552
[9]   IMPROVEMENT OF BLOOD COMPATIBILITY ON CELLULOSE DIALYSIS MEMBRANE .1. GRAFTING OF 2-METHACRYLOYLOXYETHYL PHOSPHORYLCHOLINE ON TO A CELLULOSE MEMBRANE-SURFACE [J].
ISHIHARA, K ;
NAKABAYASHI, N ;
FUKUMOTO, K ;
AOKI, J .
BIOMATERIALS, 1992, 13 (03) :145-149
[10]   PROTEIN ADSORPTION FROM HUMAN PLASMA IS REDUCED ON PHOSPHOLIPID POLYMERS [J].
ISHIHARA, K ;
ZIATS, NP ;
TIERNEY, BP ;
NAKABAYASHI, N ;
ANDERSON, JM .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (11) :1397-1407