Biomechanical analysis of the percutaneous compression plate and sliding hip screw in intracapsular hip fractures: Experimental assessment using synthetic and cadaver bones

被引:95
作者
Brandt, Erwin [1 ]
Verdonschot, Nico
van Vugt, Arie
van Kampen, Albert
机构
[1] Hosp Bernhoven, Dept Orthopaed & Trauma Surg, Oss, Netherlands
[2] Univ Med Ctr Nijmegen, Orthopaed Res Lab, Nijmegen, Netherlands
[3] Univ Med Ctr Nijmegen, Dept Trauma Surg, Nijmegen, Netherlands
[4] Univ Med Ctr Nijmegen, Dept Orthopaed Surg, Nijmegen, Netherlands
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2006年 / 37卷 / 10期
关键词
hip; fracture; biomechanics; internal fixation; intracapsular;
D O I
10.1016/j.injury.2006.05.011
中图分类号
R4 [临床医学];
学科分类号
1002 [临床医学]; 100602 [中西医结合临床];
摘要
We compared the mechanical behaviour of osteosynthesis with the percutaneous compression plate (PCCP) compared with the standard osteosynthesis sliding hip screw (SHS) in intracapsular hip fractures. We created 10 stable and 10 unstable intracapsular hip fractures in 20 synthetic femurs. Each fracture was fixed with either the SHS or PCCP. In six pairs of cadaver femurs, we created unstable intracapsular hip fractures and fixed them with the SHS or PCCP, at random on the left or right side. ALL femoral heads were exposed to a cyclic, combined axial and torque load until failure. In each group, the PCCP resisted a significantly higher toad than the SHS. Clinical prospective studies are needed to confirm these in vitro findings that the PCCP is more stable than the SHS. (C) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:979 / 983
页数:5
相关论文
共 23 条
[1]
ALBERTS KA, 1990, CLIN ORTHOP RELAT R, P129
[2]
ROTATIONAL CONTROL OF UNSTABLE SUBCAPITAL FRACTURES - EXPERIMENTAL STUDY [J].
BARIL, JD ;
SONSTEGARD, DA ;
MATTHEWS, LS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1975, 57 (05) :613-616
[3]
Booth KC, 1998, ORTHOPEDICS, V21, P1173
[4]
Percutaneous compression plating (PCCP) versus the dynamic hip screw for pertrochanteric hip fractures: preliminary results [J].
Brandt, SE ;
Lefever, S ;
Janzing, HMJ ;
Broos, PLO ;
Pilot, P ;
Houben, BJJ .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2002, 33 (05) :413-418
[5]
BRAY TJ, 1988, CLIN ORTHOP RELAT R, P127
[6]
FEMORAL-NECK FRACTURE FIXATION - COMPARISON OF A SLIDING SCREW WITH LAG SCREWS [J].
CLARK, DI ;
CROFTS, CE ;
SALEH, M .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (05) :797-800
[7]
Mechanical validation of whole bone composite femur models [J].
Cristofolini, L ;
Viceconti, M ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 1996, 29 (04) :525-535
[8]
Biomechanical comparison of internal fixation techniques for the treatment of unstable basicervical femoral neck fractures [J].
Deneka, DA ;
Simonian, PT ;
Stankewich, CJ ;
Eckert, D ;
Chapman, JR ;
Tencer, AF .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1997, 11 (05) :337-343
[9]
ENGESAETER L B, 1984, European Surgical Research, V16, P37
[10]