An investigation of liquid carryover and sample residual for a high-throughput flow cytometer sample delivery system

被引:12
作者
Bartsch, JW [1 ]
Tran, HD
Waller, A
Mammoli, AA
Buranda, T
Sklar, LA
Edwards, BS
机构
[1] Univ New Mexico, Dept Mech Engn, Albuquerque, NM 87131 USA
[2] Univ New Mexico, Ctr Hlth Sci, Canc Res & Treatment Ctr, Albuquerque, NM 87131 USA
关键词
D O I
10.1021/ac049870f
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The phenomenon of intersample contamination in air-segmented continuous-flow assays has been studied for many years, and new uses are being found for these sampling techniques every day. One application that has been developed recently employs a flow cytometer to conduct high-throughput screening assays of biological compounds. We have explored the sources of intersample contamination in the system and shown how methods developed previously can be applied to describe these phenomena. Using a simple model, we were able to accurately measure liquid film thickness in the sample tubing and demonstrate the effects of intersample contamination in a flow cytometer assay. Also, measures have been taken to reduce the level of intersample contamination in cytometric screening assays, helping to make the system a more viable tool for drug screening applications.
引用
收藏
页码:3810 / 3817
页数:8
相关论文
共 22 条
[1]   DYNAMICS OF CONTINUOUS SEGMENTED FLOW ANALYSIS [J].
BEGG, RD .
ANALYTICAL CHEMISTRY, 1971, 43 (07) :854-&
[2]   ON LIQUID FILM REMAINING IN A DRAINING CIRCULAR CYLINDRICAL VESSEL [J].
CONCUS, P .
JOURNAL OF PHYSICAL CHEMISTRY, 1970, 74 (08) :1818-&
[3]   FULLY AUTOMATED HIGH-SPEED ION-EXCHANGE CHROMATOGRAPHY OF AMINO ACIDS [J].
ERTINGSHAUSEN, G ;
ADLER, HJ ;
REICHLER, AS .
JOURNAL OF CHROMATOGRAPHY, 1969, 42 (03) :355-+
[4]  
FAIRBROTHER F, 1935, J CHEM SOC 1, P527
[5]  
Gerald CF., 1999, APPL NUMERICAL ANAL
[6]  
HAECKEL R, 1991, PURE APPL CHEM, V63, P302
[7]   LIQUID-FILM IN TAYLOR FLOW THROUGH A CAPILLARY [J].
IRANDOUST, S ;
ANDERSSON, B .
INDUSTRIAL & ENGINEERING CHEMISTRY RESEARCH, 1989, 28 (11) :1684-1688
[8]   GAS-LIQUID MASS-TRANSFER IN TAYLOR FLOW THROUGH A CAPILLARY [J].
IRANDOUST, S ;
ERTLE, S ;
ANDERSSON, B .
CANADIAN JOURNAL OF CHEMICAL ENGINEERING, 1992, 70 (01) :115-119
[9]   Mixing small volumes for continuous high-throughput flow cytometry: Performance of a mixing Y and peristaltic sample delivery [J].
Jackson, WC ;
Kuckuck, F ;
Edwards, BS ;
Mammoli, A ;
Gallegos, CM ;
Lopez, GP ;
Buranda, T ;
Sklar, LA .
CYTOMETRY, 2002, 47 (03) :183-191
[10]  
Kuckuck FW, 2001, CYTOMETRY, V44, P83, DOI 10.1002/1097-0320(20010501)44:1<83::AID-CYTO1085>3.0.CO