Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: A strategy for large-scale implementation for the treatment of carcinoma of the prostate

被引:221
作者
Burman, C [1 ]
Chui, CS [1 ]
Kutcher, G [1 ]
Leibel, S [1 ]
Zelefsky, M [1 ]
LoSasso, T [1 ]
Spirou, S [1 ]
Wu, QW [1 ]
Yang, J [1 ]
Stein, J [1 ]
Mohan, R [1 ]
Fuks, Z [1 ]
Ling, CC [1 ]
机构
[1] MEM SLOAN KETTERING CANC CTR, DEPT RADIAT ONCOL, NEW YORK, NY 10021 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1997年 / 39卷 / 04期
关键词
radiotherapy; prostate cancer; intensity modulation; dynamic multileaf collimator; DMLC;
D O I
10.1016/S0360-3016(97)00458-6
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To improve the local control of patients with adenocarcinoma of the prostate we have implemented intensity modulated radiation therapy (IMRT) to deliver a prescribed dose of 81 Gy. This method is based on inverse planning and the use of dynamic multileaf collimators (DMLC). Because IMRT is a new modality, a major emphasis was on the quality assurance of each component of the process and on patient safety. In this article we describe in detail our procedures and quality assurance program. Methods and Materials: Using an inverse algorithm, we have developed a treatment plan consisting five intensity-modulated(IM) photon fields that are delivered with DMIC. In the planning stage, the planner specifies the number of beams and their directions, and the desired doses for the target, the normal organs and the ''overlap'' regions. Then, the inverse algorithm designs intensity profiles that best meet the specified criteria. A second algorithm determines the leaf motion that would produce the designed intensity pattern and produces a DMLC file as input to the MLC control computer. Our quality assurance program for the planning and treatment delivery process includes the following components: 1) verification of the DMLC field boundary on localization pod film, 2) verification that the leaf motion of the DMLC file produces the planned dose distribution (with an independent calculation), 3) comparison of dose distribution produced by DMLC in a flat phantom with that calculated by the treatment planning computer for the same experimental condition, 4) comparison of the planned leaf motions with that implemented for the treatment (as recorded on the MLC log files), 5) confirmation of the initial and final positions of the MLC for each field by a record-and-verify system, and 6) in vivo dose measurements. Results: Using a five-field IMRT plan we have customized dose distribution to conform to and deliver 81 Gy to the PTV. In addition, in the overlap regions between the PTV and the rectum, and between the PTV and the bladder, the dose is kept within the tolerance of the respective organs. Our QA checks show acceptable agreement between the planned and the implemented leaf motions. Correspondingly, film and TLD dosimetry indicates that doses delivered agrees with the planned dose to within 2%. As of September 15, 1996, we have treated eight patients to 81 Gy with IMRT. Conclusion: For complex planning problems where the surrounding normal tissues place severe constraints on the prescription dose, IMRT provides a powerful and efficient solution. Given a comprehensive and rigorous quality-assurance program, the intensity-modulated fields can be efficaciously and accurately delivered using DMLC. IMRT treatment is now ready for routine implementation on a large scale in our clinic. (C) 1997 Elsevier Science Inc.
引用
收藏
页码:863 / 873
页数:11
相关论文
共 30 条
[1]   LATE RECTAL BLEEDING FOLLOWING COMBINED X-RAY AND PROTON HIGH-DOSE IRRADIATION FOR PATIENTS WITH STAGES T3-T4 PROSTATE CARCINOMA [J].
BENK, VA ;
ADAMS, JA ;
SHIPLEY, WU ;
URIE, MM ;
MCMANUS, PL ;
EFIRD, JT ;
WILLETT, CG ;
GOITEIN, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 26 (03) :551-557
[2]   METHODS OF IMAGE-RECONSTRUCTION FROM PROJECTIONS APPLIED TO CONFORMATION RADIOTHERAPY [J].
BORTFELD, T ;
BURKELBACH, J ;
BOESECKE, R ;
SCHLEGEL, W .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (10) :1423-1434
[3]  
BORTFELD T, 1992, ADV RAD THERAPY TUMO, P503
[4]   AUTOMATIC-GENERATION OF BEAM APERTURES [J].
BREWSTER, L ;
MAGERAS, GS ;
MOHAN, R .
MEDICAL PHYSICS, 1993, 20 (05) :1337-1342
[5]   FITTING OF NORMAL TISSUE TOLERANCE DATA TO AN ANALYTIC-FUNCTION [J].
BURMAN, C ;
KUTCHER, GJ ;
EMAMI, B ;
GOITEIN, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (01) :123-135
[6]   THE INFLUENCE OF SCATTER ON THE DESIGN OF OPTIMIZED INTENSITY MODULATIONS [J].
CHEN, Z ;
WANG, XH ;
BORTFELD, T ;
MOHAN, R ;
REINSTEIN, L .
MEDICAL PHYSICS, 1995, 22 (11) :1727-1733
[7]   OFF-CENTER RATIOS FOR 3-DIMENSIONAL DOSE CALCULATIONS [J].
CHUI, CS ;
MOHAN, R .
MEDICAL PHYSICS, 1986, 13 (03) :409-412
[8]  
GOITEIN M, 1987, REPORT WORKING GROUP
[9]   THE EFFECT OF DOSE ON LOCAL-CONTROL OF PROSTATE-CANCER [J].
HANKS, GE ;
MARTZ, KL ;
DIAMOND, JJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 15 (06) :1299-1305
[10]  
ICRU, 1993, 50 ICRU, V50, P3