Comparison of Balloon-Expandable vs Self-expandable Valves in Patients Undergoing Transcatheter Aortic Valve Replacement The CHOICE Randomized Clinical Trial

被引:561
作者
Abdel-Wahab, Mohamed [1 ]
Mehilli, Julinda [2 ]
Frerker, Christian [3 ]
Neumann, Franz-Josef [4 ]
Kurz, Thomas [5 ]
Toelg, Ralph [1 ]
Zachow, Dirk [6 ]
Guerra, Elena [7 ]
Massberg, Steffen [2 ]
Schaefer, Ulrich [3 ,8 ]
El-Mawardy, Mohamed [1 ]
Richardt, Gert [1 ]
机构
[1] Segeberger Kliniken, Ctr Heart, Bad Segeberg, Germany
[2] Munich Univ Clin, Munich, Germany
[3] Asklepios Clin St Georg, Dept Cardiol, Hamburg, Germany
[4] Univ Heart Ctr Freiburg Bad Krozingen, Bad Krozingen, Germany
[5] Lubeck Univ Hosp, Dept Cardiol, Lubeck, Germany
[6] Segeberger Kliniken, Dept Radiol, Bad Segeberg, Germany
[7] Intracoronary Stenting & Antithrombosis Res Ctr, Munich, Germany
[8] Univ Hosp Hamburg Eppendorf, Dept Cardiol, Hamburg, Germany
来源
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION | 2014年 / 311卷 / 15期
关键词
EDWARDS-SAPIEN; IMPLANTATION INCIDENCE; MEDTRONIC-COREVALVE; REGURGITATION; STENOSIS; OUTCOMES; PREDICTORS; ROOT;
D O I
10.1001/jama.2014.3316
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
IMPORTANCE Transcatheter aortic valve replacement (TAVR) is an effective treatment option for high-risk patients with severe aortic stenosis. Different from surgery, transcatheter deployment of valves requires either a balloon-expandable or self-expandable system. A randomized comparison of these 2 systems has not been performed. OBJECTIVE To determine whether the balloon-expandable device is associated with a better success rate than the self-expandable device. DESIGN, SETTING, AND PATIENTS The CHOICE study was an investigator-initiated trial in high-risk patients with severe aortic stenosis and an anatomy suitable for the transfemoral TAVR procedure. One hundred twenty-one patients were randomly assigned to receive a balloon-expandable valve (Edwards Sapien XT) and 120 were assigned to receive a self-expandable valve (Medtronic CoreValve). Patients were enrolled between March 2012 and December 2013 at 5 centers in Germany. INTERVENTIONS Transfemoral TAVR with a balloon-expandable or self-expandable device. MAIN OUTCOMES AND MEASURES The primary end point was device success, which is a composite end point including successful vascular access and deployment of the device and retrieval of the delivery system, correct position of the device, intended performance of the heart valve without moderate or severe regurgitation, and only 1 valve implanted in the proper anatomical location. Secondary end points included cardiovascular mortality, bleeding and vascular complications, postprocedural pacemaker placement, and a combined safety end point at 30 days, including all-cause mortality, major stroke, and other serious complications. RESULTS Device success occurred in 116 of 121 patients (95.9%) in the balloon-expandable valve group and 93 of 120 patients (77.5%) in the self-expandable valve group (relative risk [RR], 1.24, 95% CI, 1.12-1.37, P <.001). This was attributed to a significantly lower frequency of residual more-than-mild aortic regurgitation (4.1% vs 18.3%; RR, 0.23; 95% CI, 0.09-0.58; P <.001) and the less frequent need for implanting more than 1 valve (0.8% vs 5.8%, P = .03) in the balloon-expandable valve group. Cardiovascular mortality at 30 days was 4.1% in the balloon-expandable valve group and 4.3% in the self-expandable valve group (RR, 0.97; 95% CI, 0.29-3.25; P = .99). Bleeding and vascular complications were not significantly different, and the combined safety end point occurred in 18.2% of those in the balloon-expandable valve group and 23.1% of the self-expandable valve group (RR, 0.79; 95% CI, 0.48-1.30; P = .42). Placement of a new permanent pacemaker was less frequent in the balloon-expandable valve group (17.3% vs 37.6%, P = .001). CONCLUSIONS AND RELEVANCE Among patients with high-risk aortic stenosis undergoing TAVR, the use of a balloon-expandable valve resulted in a greater rate of device success than use of a self-expandable valve.
引用
收藏
页码:1503 / 1514
页数:12
相关论文
共 35 条
[1]
Aortic Regurgitation After Transcatheter Aortic Valve Implantation With Balloon- and Self-Expandable Prostheses A Pooled Analysis From a 2-Center Experience [J].
Abdel-Wahab, Mohamed ;
Comberg, Thomas ;
Buettner, Heinz Joachim ;
El-Mawardy, Mohamed ;
Chatani, Kenichi ;
Gick, Michael ;
Geist, Volker ;
Richardt, Gert ;
Neumann, Franz-Josef .
JACC-CARDIOVASCULAR INTERVENTIONS, 2014, 7 (03) :284-292
[2]
Aortic regurgitation after transcatheter aortic valve implantation: incidence and early outcome. Results from the German transcatheter aortic valve interventions registry [J].
Abdel-Wahab, Mohamed ;
Zahn, Ralf ;
Horack, Martin ;
Gerckens, Ulrich ;
Schuler, Gerhard ;
Sievert, Horst ;
Eggebrecht, Holger ;
Senges, Jochen ;
Richardt, Gert .
HEART, 2011, 97 (11) :899-906
[3]
Incidence, Predictors, and Outcomes of Aortic Regurgitation After Transcatheter Aortic Valve Replacement Meta-Analysis and Systematic Review of Literature [J].
Athappan, Ganesh ;
Patvardhan, Eshan ;
Tuzcu, E. Murat ;
Svensson, Lars Georg ;
Lemos, Pedro A. ;
Fraccaro, Chiara ;
Tarantini, Giuseppe ;
Sinning, Jan-Malte ;
Nickenig, Georg ;
Capodanno, Davide ;
Tamburino, Corrado ;
Latib, Azeem ;
Colombo, Antonio ;
Kapadia, Samir R. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 61 (15) :1585-1595
[4]
Anatomical and Procedural Features Associated With Aortic Root Rupture During Balloon-Expandable Transcatheter Aortic Valve Replacement [J].
Barbanti, Marco ;
Yang, Tae-Hyun ;
Cabau, Josep Rodes ;
Tamburino, Corrado ;
Wood, David A. ;
Jilaihawi, Hasan ;
Blanke, Phillip ;
Makkar, Raj R. ;
Latib, Azeem ;
Colombo, Antonio ;
Tarantini, Giuseppe ;
Raju, Rekha ;
Binder, Ronald K. ;
Nguyen, Giang ;
Freeman, Melanie ;
Ribeiro, Henrique B. ;
Kapadia, Samir ;
Min, James ;
Feuchtner, Gudrun ;
Gurtvich, Ronen ;
Alqoofi, Faisal ;
Pelletier, Marc ;
Ussia, Gian Paolo ;
Napodano, Massimo ;
de Brito, Fabio Sandoli ;
Kodali, Susheel ;
Norgaard, Bjarne L. ;
Hansson, Nicolaj C. ;
Pache, Gregor ;
Canovas, Sergio J. ;
Zhang, Hongbin ;
Leon, Martin B. ;
Webb, John G. ;
Leipsic, Jonathon .
CIRCULATION, 2013, 128 (03) :244-253
[5]
Transcatheter Aortic Valve Implantation With the Edwards SAPIEN Versus the Medtronic CoreValve Revalving System Devices A Multicenter Collaborative Study: The PRAGMATIC Plus Initiative (Pooled-RotterdAm-Milano-Toulouse In Collaboration) [J].
Chieffo, Alaide ;
Buchanan, Gill Louise ;
Van Mieghem, Nicolas M. ;
Tchetche, Didier ;
Dumonteil, Nicolas ;
Latib, Azeem ;
van der Boon, Robert M. A. ;
Vahdat, Olivier ;
Marcheix, Bertrand ;
Farah, Bruno ;
Serruys, Patrick W. ;
Fajadet, Jean ;
Carrie, Didier ;
de Jaegere, Peter P. T. ;
Colombo, Antonio .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 61 (08) :830-836
[6]
Paravalvular Leak After Transcatheter Aortic Valve Replacement The New Achilles' Heel? A Comprehensive Review of the Literature [J].
Genereux, Philippe ;
Head, Stuart J. ;
Hahn, Rebecca ;
Daneault, Benoit ;
Kodali, Susheel ;
Williams, Mathew R. ;
van Mieghem, Nicolas M. ;
Alu, Maria C. ;
Serruys, Patrick W. ;
Kappetein, A. Pieter ;
Leon, Martin B. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 61 (11) :1125-1136
[7]
Registry of Transcatheter Aortic-Valve Implantation in High-Risk Patients [J].
Gilard, Martine ;
Eltchaninoff, Helene ;
Iung, Bernard ;
Donzeau-Gouge, Patrick ;
Chevreul, Karine ;
Fajadet, Jean ;
Leprince, Pascal ;
Leguerrier, Alain ;
Lievre, Michel ;
Prat, Alain ;
Teiger, Emmanuel ;
Lefevre, Thierry ;
Himbert, Dominique ;
Tchetche, Didier ;
Carrie, Didier ;
Albat, Bernard ;
Cribier, Alain ;
Rioufol, Gilles ;
Sudre, Arnaud ;
Blanchard, Didier ;
Collet, Frederic ;
Dos Santos, Pierre ;
Meneveau, Nicolas ;
Tirouvanziam, Ashok ;
Caussin, Christophe ;
Guyon, Philippe ;
Boschat, Jacques ;
Le Breton, Herve ;
Collart, Frederic ;
Houel, Remi ;
Delpine, Stephane ;
Souteyrand, Geraud ;
Favereau, Xavier ;
Ohlmann, Patrick ;
Doisy, Vincent ;
Grollier, Gilles ;
Gommeaux, Antoine ;
Claudel, Jean-Philippe ;
Bourlon, Francois ;
Bertrand, Bernard ;
Van Belle, Eric ;
Laskar, Marc .
NEW ENGLAND JOURNAL OF MEDICINE, 2012, 366 (18) :1705-1715
[8]
Percutaneous aortic valve replacement for severe aortic stenosis in high-rick patients using the second- and current third-generation self-expanding CoreValve prosthesis - Device success and 30-day clinical outcome [J].
Grube, Eberhard ;
Schuler, Gerhard ;
Buellesfeld, Lutz ;
Gerckens, Ulrich ;
Linke, Axel ;
Wenaweser, Peter ;
Sauren, Barthel ;
Mohr, Friedrich-Wilhelm ;
Walther, Thomas ;
Zickmann, Bernfried ;
Iversen, Stein ;
Felderhoff, Thomas ;
Cartier, Raymond ;
Bonan, Raoul .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 50 (01) :69-76
[9]
Progress and Current Status of Percutaneous Aortic Valve Replacement: Results of Three Device Generations of the CoreValve Revalving System [J].
Grube, Eberhard ;
Buellesfeld, Lutz ;
Mueller, Ralf ;
Sauren, Barthel ;
Zickmann, Bernfried ;
Nair, Dinesh ;
Beucher, Harald ;
Felderhoff, Thomas ;
Iversen, Stein ;
Gerckens, Ulrich .
CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2008, 1 (03) :167-175
[10]
Standardized Imaging for Aortic Annular Sizing Implications for Transcatheter Valve Selection [J].
Kasel, Albert M. ;
Cassese, Salvatore ;
Bleiziffer, Sabine ;
Amaki, Makoto ;
Hahn, Rebecca T. ;
Kastrati, Adnan ;
Sengupta, Partho P. .
JACC-CARDIOVASCULAR IMAGING, 2013, 6 (02) :249-262