Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room

被引:61
作者
Dumas, C [1 ]
Wahr, JA [1 ]
Tremper, KK [1 ]
机构
[1] UNIV MICHIGAN,DEPT ANESTHESIOL,ANN ARBOR,MI 48109
关键词
D O I
10.1097/00000539-199608000-00012
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The frequency and nature of spurious pulse oximetry readings were compared using both a conventional pulse oximeter (CPO) and a prototype Masimo signal extraction technology pulse oximeter (Masimo SET(TM)). At a university hospital, 50 ASA physical status I-IV adult patients who underwent general or spinal-epidural anesthesia were selected from a group of 250 patients on the basis of high-alarm generation with routine postoperative pulse oximetry. Pulse oximetry data were recorded simultaneously from both devices with a computer. Overall, the CPO alarm frequency (i.e., oxygen saturation <90%, or complete signal loss) was once every 13 min, and 87% of these alarms were considered false. Alarms were considered false based on reference electrocardiographs (16 patients), arterial blood gases (7 patients), and clinical assessment. The prototype Masimo SET(TM) device alarm frequency was once every 30 min, and 59% of these were considered false. During arm motion with 15 patients, the CPO device produced spurious signals on 54 occasions compared with five for prototype Masimo SET(TM). The incidence of artifactual pulse oximetry events during patient motion appear to be substantially reduced with the prototype Masimo SET(TM) device, relative to a CPO device.
引用
收藏
页码:269 / 272
页数:4
相关论文
共 16 条
  • [1] BARKER SJ, 1995, ANESTHESIOLOGY, V83, pA450
  • [2] EICHHORN JH, 1986, JAMA-J AM MED ASSOC, V256, P1017
  • [3] BLOOD-FLOW LIMITS AND PULSE OXIMETER SIGNAL-DETECTION
    LAWSON, D
    NORLEY, I
    KORBON, G
    LOEB, R
    ELLIS, J
    [J]. ANESTHESIOLOGY, 1987, 67 (04) : 599 - 603
  • [4] PAN PH, 1991, ANESTHESIOLOGY, V75, pA405
  • [5] RIAZI J, 1993, ANESTHESIOLOGY, V79, pA530
  • [6] PULSE OXIMETER FAILURE THRESHOLDS IN HYPOTENSION AND VASOCONSTRICTION
    SEVERINGHAUS, JW
    SPELLMAN, MJ
    [J]. ANESTHESIOLOGY, 1990, 73 (03) : 532 - 537
  • [7] SEVERINGHAUS JW, 1992, ANESTHESIOLOGY, V76, P1018
  • [8] SHAH N, 1995, ANESTHESIOLOGY, V83, pA1067
  • [9] SHAH N, 1995, ANESTHESIOLOGY, V83, pA452
  • [10] SHAH NK, 1993, ANESTHESIOLOGY, V79, pA531