First MR images obtained during megavoltage photon irradiation from a prototype integrated linac-MR system

被引:269
作者
Fallone, B. G. [1 ,2 ]
Murray, B. [1 ,2 ]
Rathee, S. [1 ,2 ]
Stanescu, T. [1 ]
Steciw, S. [1 ,2 ]
Vidakovic, S. [1 ,2 ]
Blosser, E. [1 ]
Tymofichuk, D. [1 ]
机构
[1] Cross Canc Inst, Dept Med Phys, Edmonton, AB T6G 1Z2, Canada
[2] Univ Alberta, Dept Oncol, Med Phys Div, Edmonton, AB T6G 1Z2, Canada
关键词
biomedical MRI; cancer; linear accelerators; phantoms; radiation therapy; 0.2; T;
D O I
10.1118/1.3125662
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The authors report the first magnetic resonance (MR) images produced by their prototype MR system integrated with a radiation therapy source. The prototype consists of a 6 MV linac mounted onto the open end of a biplanar 0.2 T permanent MR system which has 27.9 cm pole-to-pole opening with flat gradients (40 mT/m) running under a TMX NRC console. The distance from the magnet isocenter to the linac target is 80 cm. The authors' design has resolved the mutual interferences between the two devices such that the MR magnetic field does not interfere with the trajectory of the electron in the linac waveguide, and the radiofrequency (RF) signals from each system do not interfere with the operation of the other system. Magnetic and RF shielding calculations were performed and confirmed with appropriate measurements. The prototype is currently on a fixed gantry; however, in the very near future, the linac and MR magnet will rotate in unison such that the linac is always aimed through the opening in the biplanar magnet. MR imaging was found to be fully operational during linac irradiation and proven by imaging a phantom with conventional gradient echo sequences. Except for small changes in SNR, MR images produced during irradiation were visually and quantitatively very similar to those taken with the linac turned off. This prototype system provides proof of concept that the design has decreased the mutual interferences sufficiently to allow the development of real-time MR-guided radiotherapy. Low field-strength systems (0.2-0.5 T) have been used clinically as diagnostic tools. The task of the linac-MR system is, however, to provide MR guidance to the radiotherapy beam. Therefore, the 0.2 T field strength would provide adequate image quality for this purpose and, with the addition of fast imaging techniques, has the potential to provide 4D soft-tissue visualization not presently available in image-guided radiotherapy systems. The authors' initial design incorporates a permanent magnet; however, other types of magnets and field strengths could also be incorporated. Usable MR images were obtained during linac irradiation from the linac-MR prototype. The authors' prototype design can be used as the functional starting point in developing real-time MR guidance offering soft-tissue contrast that can be coupled with tumor tracking for real-time adaptive radiotherapy.
引用
收藏
页码:2084 / 2088
页数:5
相关论文
共 12 条
[1]   MR-guided radiofrequency ablation in a 0.2-T open MR system: Technical success and technique effectiveness in 100 liver tumors [J].
Clasen, Stephan ;
Boss, Andreas ;
Schmidt, Diethard ;
Schraml, Christina ;
Fritz, Jan ;
Schick, Fritz ;
Claussen, Claus D. ;
Pereira, Philippe L. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 26 (04) :1043-1052
[2]   A device for realtime 3D image-guided IMRT [J].
Dempsey, JF ;
Benoit, D ;
Fitzsimmons, JR ;
Haghighat, A ;
Li, JG ;
Low, DA ;
Mutic, S ;
Palta, JR ;
Romeijn, HE ;
Sjoden, GE .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02) :S202-S202
[3]   Development of a Linac-MRI system for real-time ART [J].
Fallone, B. ;
Carlone, M. ;
Murray, B. ;
Rathee, S. ;
Stanescu, T. ;
Steciw, S. ;
Wachowicz, K. ;
Kirkby, C. .
MEDICAL PHYSICS, 2007, 34 (06) :2547-2547
[4]   Low-field MRI of the ankle joint-first experience in the pediatric age group. [J].
Herber, S ;
Kreitner, KF ;
Kalden, P ;
Löw, R ;
Berger, S ;
Thelen, M .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (03) :267-273
[5]   Line-scan diffusion tensor MR imaging at 0.2 T: Feasibility study [J].
Hori, M ;
Aoki, S ;
Okubo, T ;
Ishigame, K ;
Kumagai, H ;
Araki, T .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 22 (06) :794-798
[6]   Patient dosimetry for hybrid MRI-radiotherapy systems [J].
Kirkby, C. ;
Stanescu, T. ;
Rathee, S. ;
Carlone, M. ;
Murray, B. ;
Fallone, B. G. .
MEDICAL PHYSICS, 2008, 35 (03) :1019-1027
[7]   In room magnetic resonance imaging guided radiotherapy (MRIgRT) [J].
Lagendijk, J ;
Raaymakers, B ;
van der Heide, U ;
Overweg, J ;
Brown, K ;
Bakker, C ;
Raaijmakers, A ;
Vulpen, M ;
Welleweerd, J ;
Jürgenliemk-Schulz, I .
MEDICAL PHYSICS, 2005, 32 (06) :2067-2067
[8]   MR arthrography of the shoulder: comparison of low-field (0.2 T) vs high-field (1.5 T) imaging [J].
Loew, R ;
Kreitner, KF ;
Runkel, M ;
Zoellner, J ;
Thelen, M .
EUROPEAN RADIOLOGY, 2000, 10 (06) :989-996
[9]   MR lung imaging at 0.2 T with T1-weighted true FISP:: Native and oxygen-enhanced [J].
Müller, CJ ;
Löffler, R ;
Deimling, M ;
Peller, M ;
Reiser, M .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (02) :164-168
[10]   Magnetic-field-induced dose effects in MR-guided radiotherapy systems: dependence on the magnetic field strength [J].
Raaijmakers, A. J. E. ;
Raaymakers, B. W. ;
Lagendijk, J. J. W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (04) :909-923