Isotope-ratio-monitoring liquid chromatography mass spectrometry (IRM-LCMS): First results from a moving wire interface system

被引:32
作者
Brand, WA [1 ]
Dobberstein, P [1 ]
机构
[1] FINNIGAN MAT GMBH,D-28197 BREMEN,GERMANY
关键词
carbon; 13; isotope ratios; liquid chromatography; mass spectrometry; on-line coupling;
D O I
10.1080/10256019608036320
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A Liquid Chromatography-Combustion (LC-C) Interface, based on a moving wire technique, has been built and tested. The LC effluent is deposited onto a transport wire, which carries the sample through solvent evaporation and combustion ovens. CO2 from the combustion step is analysed in an isotope ratio mass spectrometer. Performance of the interface was tested by loop injections of sucrose and glucose into a liquid flow of methanol/water (80/20). Accuracy and precision of delta(13)C(PDB)<1 parts per thousand was achieved for sample concentrations >500 mu g/ul (5 mu l loop), sufficient for studies at natural isotope ratios. In case of C-13 tracer applications the detection limit was determined to be about 20 pg carbon tracer (on wire).