Problems associated with the 25-gauge transconjunctival sutureless vitrectomy system during and after surgery

被引:71
作者
Byeon, Suk Ho
Chu, Young Kwang
Lee, Sung Chul
Koh, Hyoung Jun
Kim, Sung Soo
Kwon, Oh Woong
机构
[1] Yonsei Univ, Coll Med, Dept Ophthalmol, Inst Vis Res, Seoul 120752, South Korea
[2] Yonsei Univ, Coll Med, Youndong Severance Hosp, Seoul 120752, South Korea
[3] Siloam Eye Hosp, Seoul, South Korea
关键词
transconjunctival sutureless vitrectomy; 25-gauge vitrectomy; conventional vitrectomy; microcannula; hypotony; lens damage; retinal detachment; glaucoma; choroidal detachment;
D O I
10.1159/000093081
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To report the initial experiences with the 25-gauge transconjunctival sutureless vitrectomy (TSV) system, including intraoperative and postoperative problems. Methods: We retrospectively reviewed the medical records and surgical videotapes of 50 consecutive patients who underwent vitrectomy performed by one surgeon using the TSV system. Results: Intraoperatively, we encountered such problems as difficulty in inserting the microcannula, which led to deformity, instability of the microcannula, self-disconnection of the infusion tip and resultant lens damage, and conversion to 20-gauge conventional vitrectomy. Postoperatively, there were 8 cases with hypotony (IOP < 6 mm Hg) on day 1, 6 cases with elevated IOP, and 3 cases with retinal detachment during follow-up. Conclusion: Though certain problems exist during and after surgeries using TSV, this system is both convenient and safe for various vitreoretinal procedures. Copyright (c) 2006 S. Karger AG, Basel.
引用
收藏
页码:259 / 265
页数:7
相关论文
共 19 条
[2]   Sutureless pars plana vitrectomy through self-sealing sclerotomies [J].
Chen, JC .
ARCHIVES OF OPHTHALMOLOGY, 1996, 114 (10) :1273-1275
[3]   Vitreoretinal surgery using Transconjunctival Sutureless Vitrectomy [J].
Cho, YJ ;
Lee, JM ;
Kim, SS .
YONSEI MEDICAL JOURNAL, 2004, 45 (04) :615-620
[4]   Initial experience using the Transconjunctival Sutureless Vitrectomy System for vitreoretinal surgery [J].
Fujii, GY ;
de Juan, E ;
Humayun, MS ;
Chang, TS ;
Pieramici, DJ ;
Barnes, A ;
Kent, D .
OPHTHALMOLOGY, 2002, 109 (10) :1814-1820
[5]   A new 25-gauge instrument system for transconjunctival sutureless vitrectomy surgery [J].
Fujii, GY ;
de Juan, E ;
Humayun, MS ;
Pieramici, DJ ;
Chang, TS ;
Ng, E ;
Barnes, A ;
Wu, SL ;
Sommerville, DN .
OPHTHALMOLOGY, 2002, 109 (10) :1807-1812
[6]  
HAN DP, 1989, OPHTHALMOLOGY, V96, P1357
[7]   Office-based sutureless transconjunctival pars plana vitrectomy [J].
Hilton, GF ;
Josephberg, RG ;
Halperin, LS ;
Madreperla, SA ;
Brinton, DA ;
Lee, SS ;
Gordon, SF .
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2002, 22 (06) :725-732
[8]   Longer-term outcomes of transconjunctival sutureless 25-gauge vitrectomy [J].
Ibarra, MS ;
Hermel, M ;
Prenner, JL ;
Hassan, TS .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2005, 139 (05) :831-836
[9]   Intraoperative breakage of a 25-gauge vitreous cutter [J].
Inoue, M ;
Noda, K ;
Ishida, S ;
Nagai, N ;
Imamura, Y ;
Oguchi, Y .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2004, 138 (05) :867-869
[10]   Modified sutureless sclerotomies in pars plana vitrectomy [J].
Jackson, T .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2000, 129 (01) :116-117