Radiometric measurement comparisons using transfer radiometers in support of the calibration of NASA's Earth Observing System (EOS) sensors

被引:17
作者
Butler, JJ [1 ]
Johnson, BC [1 ]
Brown, SW [1 ]
Yoon, HW [1 ]
Barnes, RA [1 ]
Markham, BL [1 ]
Biggar, SF [1 ]
Zalewski, EF [1 ]
Spyak, PR [1 ]
Cooper, JW [1 ]
Sakuma, F [1 ]
机构
[1] NASA, Goddard Space Flight Ctr, Greenbelt, MD 20771 USA
来源
SENSORS, SYSTEMS, AND NEXT-GENERATION SATELLITES III | 1999年 / 3870卷
关键词
calibration; integrating spheres; radiance; transfer radiometers;
D O I
10.1117/12.373186
中图分类号
V [航空、航天];
学科分类号
08 ; 0825 ;
摘要
EOS satellite instruments operating in the visible through the shortwave infrared wavelength regions (from 0.4 mu m to 2.5 mu m) are calibrated prior to flight for radiance response using integrating spheres at a number of instrument builder facilities. The traceability of the radiance produced by these spheres with respect to international standards is the responsibility of the instrument builder, and different calibration techniques are employed by those builders. The National Aeronautics and Space Administration's (NASA's) Earth Observing System (EOS) Project Science Office, realizing the importance of preflight calibration and cross-calibration, has sponsored a number of radiometric measurement comparisons, the main purpose of which is to validate the radiometric scale assigned to the integrating spheres by the instrument builders. This paper describes the radiometric measurement comparisons, the use of stable transfer radiometers to perform the measurements, and the measurement approaches and protocols used to validate integrating sphere radiances. Stable transfer radiometers from the National Institute of Standards and Technology, the University of Arizona Optical Sciences Center Remote Sensing Group, NASA's Goddard Space Flight Center, and the National Research Laboratory of Metrology in Japan, have participated in these comparisons. The approaches used in the comparisons include the measurement of multiple integrating sphere lamp levels, repeat measurements of select lamp levels, the use of the stable radiometers as external sphere monitors, and the rapid reporting of measurement results. Results from several comparisons are presented. The absolute radiometric calibration standard uncertainties required by the EOS satellite instruments are typically in the +/-3% to +/-5% range. Preliminary results reported during eleven radiometric measurement comparisons held between February 1995 and May 1998 have shown the radiance of integrating spheres agreed to within +/-2.5% from the average at blue wavelengths and to within +/-1.7% from the average at red and near infrared wavelengths. This level of agreement lends confidence in the use of the transfer radiometers in validating the radiance scales assigned by EOS instrument calibration facilities to their integrating sphere sources.
引用
收藏
页码:180 / 192
页数:13
相关论文
共 9 条
[1]  
[Anonymous], EARTH OBSERVER
[2]  
BIGGAR SF, 1993, P SOC PHOTO-OPT INS, V1939, P243, DOI 10.1117/12.152851
[3]   Radiometric measurement comparison using the Ocean Color Temperature Scanner (OCTS) visible and near infrared integrating sphere [J].
Johnson, BC ;
Sakuma, F ;
Butler, JJ ;
Biggar, SF ;
Cooper, JW ;
Ishida, J ;
Suzuki, K .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1997, 102 (06) :627-646
[4]  
JOHNSON BC, 1998, TM1998206892 NASA
[5]  
JOHNSON BC, 1996, 104566 NASA TM, V37
[6]   Monitoring large-aperture spherical integrating sources with a portable radiometer during satellite instrument calibration [J].
Markham, BL ;
Schafer, JS ;
Wood, FM ;
Dabney, PW ;
Barker, JL .
METROLOGIA, 1998, 35 (04) :643-648
[7]  
SAKUMA F, 1995, P SOC PHOTO-OPT INS, V2553, P232, DOI 10.1117/12.221367
[8]  
SAKUMA F, 1994, P IGARSS
[9]   A 400 nm to 2500 nm absolute spectral radiance comparison using filter radiometers [J].
Yoon, HW ;
Johnson, BC ;
Kelch, D ;
Biggar, S ;
Spyak, PR .
METROLOGIA, 1998, 35 (04) :563-568