Influence of anatomical, physical, and detection-system parameters on surface EMG

被引:261
作者
Farina, D
Cescon, C
Merletti, R
机构
[1] Politecn Torino, Ctr Bioingn, Dept Elect, I-10129 Turin, Italy
[2] Ecole Cent Nantes, Dept Automat & Informat Appl, Nantes, France
关键词
D O I
10.1007/s00422-002-0309-2
中图分类号
TP3 [计算技术、计算机技术];
学科分类号
0812 ;
摘要
Many previous studies were focused on the influence of anatomical, physical, and detection-system parameters on recorded surface EMG signals. Most of them were conducted by simulations. Previous EMG models have been limited by simplifications which did not allow simulation of several aspects of the EMG generation and detection systems. We recently proposed a model for fast and accurate simulation of the surface EMG. It characterizes the volume conductor as a non-homogeneous and anisotropic medium, and allows simulation of EMG signals generated by finite-length fibers without approximation of the current-density source. The influence of thickness of the subcutaneous tissue layers, fiber inclination, fiber depth, electrode size and shape, spatial filter transfer function, interelectrode distance, length of the fibers on surface, single-fiber action-potential amplitude, frequency content, and estimated conduction velocity are investigated in this paper. Implications of the results on electrode positioning procedures, spatial filter design, and EMG signal interpretation are discussed.
引用
收藏
页码:445 / 456
页数:12
相关论文
共 59 条
[1]   SELECTIVE NONINVASIVE ELECTRODE TO STUDY MYOELECTRIC SIGNALS [J].
BHULLAR, HK ;
LOUDON, GH ;
FOTHERGILL, JC ;
JONES, NB .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1990, 28 (06) :581-586
[2]  
BLANC Y, 2000, P 12 JOURN FRANC EL, P39
[3]  
BLOK JH, 1997, SENIAM DELIVERABLE, V5, P60
[4]   SURFACE MYOELECTRIC SIGNAL CROSS-TALK AMONG MUSCLES OF THE LEG [J].
DE LUCA, CJ ;
MERLETTI, R .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1988, 69 (06) :568-575
[5]   Precise and fast calculation of the motor unit potentials detected by a point and rectangular plate electrode [J].
Dimitrov, GV ;
Dimitrova, NA .
MEDICAL ENGINEERING & PHYSICS, 1998, 20 (05) :374-381
[6]  
DIMITROVA N, 1974, Electromyography and Clinical Neurophysiology, V14, P53
[7]   Calculation of extracellular potentials produced by an inclined muscle fibre at a rectangular plate electrode [J].
Dimitrova, NA ;
Dimitrov, AG ;
Dimitrov, GV .
MEDICAL ENGINEERING & PHYSICS, 1999, 21 (08) :583-588
[8]   Effect of electrode dimensions on motor unit potentials [J].
Dimitrova, NA ;
Dimitrov, GV ;
Chihman, VN .
MEDICAL ENGINEERING & PHYSICS, 1999, 21 (6-7) :479-485
[9]   Improvement of spatial resolution in surface-EMG: A theoretical and experimental comparison of different spatial filters [J].
DisselhorstKlug, C ;
Silny, J ;
Rau, G .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1997, 44 (07) :567-574
[10]  
DISSELHORSTKLUG C, 1999, SENIAM DELIVERABLE, V4, P38