Six-year experience of caring for forty-four patients with a left ventricular assist device at home: Safe, economical, necessary

被引:50
作者
Morales, DLS [1 ]
Catanese, KA [1 ]
Helman, DN [1 ]
Williams, MR [1 ]
Weinberg, A [1 ]
Goldstein, DJ [1 ]
Rose, EA [1 ]
Oz, MC [1 ]
机构
[1] Columbia Univ Coll Phys & Surg, Dept Surg, New York, NY 10032 USA
关键词
D O I
10.1016/S0022-5223(00)70180-2
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
Objective:With increasing numbers of implantations, left ventricular assist device programs can put a financial strain on a hospital unless an efficient and safe outpatient program is developed. However, the left ventricular assist device is not widely recognized in the medical community as being reliable enough to support a patient at home. We reviewed our experience with these patients at home to assess the safety and the benefits of such a program, Methods: Our institutional 6-year experience with 90 consecutive recipients of a wearable left ventricular assist device was analyzed. Results: Forty-four (49%) of the 90 patients who received TCI vented-electric left ventricular assist devices (Thermo Cardio-systems, Inc, Woburn, Mass) were discharged, spending a total of 4546 days (12.5 years) at home with an average of 103 +/- 16 days of outpatient support (range 9-436 days). Of these 44 patients, all were successfully bridged to, transplantation (42 patients, 96%) or planned explantation (2 patients, 4%), None of the outpatients died. The cumulative events per outpatient month were 0.020 for bleeding, 0.053 for device infection, 0.0068 for thromboembolus, and 0.020 for major malfunctions. Our estimated average cost to bridge a patient to transplantation or explantation once discharged is $13,200 and as an inpatient over the same length of time, including only room and board, is $165,200, Thirty percent of outpatients were able to return to work or school, 33% to sexual activity, and 44% to driving. All outpatients performed activities of daily living, Conclusion: Current left ventricular assist device technology provides effective and economical outpatient support and is associated with limited morbidity and a satisfactory quality of life.
引用
收藏
页码:251 / 258
页数:8
相关论文
共 26 条
[1]
Cost analysis of the Novacor Left Ventricular Assist System as an outpatient bridge to heart transplantation [J].
Arabia, FA ;
Smith, RG ;
Jaffe, C ;
Wild, JC ;
Rose, DS ;
Nelson, RJ ;
McClellan, DM ;
Acuna, GA ;
Edling, NG ;
Harrington, NK ;
Rabago, G ;
Tsen, AC ;
Arzouman, DA ;
Sethi, GK ;
Copeland, JG .
ASAIO JOURNAL, 1996, 42 (05) :M546-M549
[2]
Argenziano M, 1997, J HEART LUNG TRANSPL, V16, P822
[3]
Outpatient left ventricular assist device support: A destination rather than a bridge [J].
Catanese, KA ;
Goldstein, DJ ;
Williams, DL ;
Foray, AT ;
Illick, CD ;
Gardocki, MT ;
Weinberg, AD ;
Levin, HR ;
Rose, EA ;
Oz, MC .
ANNALS OF THORACIC SURGERY, 1996, 62 (03) :646-652
[4]
Implantable left ventricular assist devices provide an excellent outpatient bridge to transplantation and recovery [J].
DeRose, JJ ;
Umana, JP ;
Argenziano, M ;
Catanese, KA ;
Gardocki, MT ;
Flannery, M ;
Levin, HR ;
Sun, BC ;
Rose, EA ;
Oz, MC .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 30 (07) :1773-1777
[5]
Implantable left ventricular assist devices - An evolving long-term cardiac replacement therapy [J].
DeRose, JJ ;
Argenziano, M ;
Sun, BC ;
Reemtsma, K ;
Oz, MC ;
Rose, EA .
ANNALS OF SURGERY, 1997, 226 (04) :461-468
[6]
Dew M A, 1993, ASAIO J, V39, P145, DOI 10.1097/00002480-199304000-00014
[7]
Quality of life outcomes in left ventricular assist system inpatients and outpatients [J].
Dew, MA ;
Kormos, RL ;
Winowich, S ;
Nastala, CJ ;
Borovetz, HS ;
Roth, LH ;
Sanchez, J ;
Griffith, BP .
ASAIO JOURNAL, 1999, 45 (03) :218-225
[8]
FEY O, 1997, EUR J CARDIO-THORAC, V11, P851
[10]
FIRST USE OF AN UNTETHERED, VENTED ELECTRIC LEFT-VENTRICULAR ASSIST DEVICE FOR LONG-TERM SUPPORT [J].
FRAZIER, OH .
CIRCULATION, 1994, 89 (06) :2908-2914