The validity of a new noncontact tonometer and its comparison with the goldmann tonometer

被引:20
作者
Lam, AYC [1 ]
Chan, RF [1 ]
Chiu, R [1 ]
Lam, CH [1 ]
机构
[1] Hong Kong Polytech Univ, Dept Optometry & Radiog, Hong Kong, Hong Kong, Peoples R China
关键词
intraocular pressure; Goldmann tonometry; noncontact tonometry;
D O I
10.1097/01.opx.0000141796.95597.43
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. This study compared a new noncontact tonometer (NCT), the Nidek NT-4000 (Nidek Co., Ltd., Aichi, Japan), with a pulse detection feature; an earlier model, the Nidek NT-2000, without this pulse detection facility; and Goldmann tonometry. Methods. Thirty-one young subjects had their intraocular pressure (IOP) in one eye randomly measured by two NCT's, followed by a Goldmann tonometer. The pulse detection mode used by the NT-4000 included synchronizing the IOP with the peak of the pulse (P-P), the middle of the pulse (P-M), or the bottom of the pulse (P-B). The order of measurements from these three modes was also random. Three consecutive readings were obtained from each mode/tonometer, and the mean was used for analysis. Results. Goldmann tonometry demonstrated the smallest variation from consecutive measurements (coefficient of variation, <4%), followed by the NT-4000 (coefficient of variation, <5% in various modes) and the NT-2000 (coefficient of variation, >5%). For the NT-4000, the mean IOP decreased from the highest in the P-P mode (16.1 mm Hg) to the lowest in the P-B mode (14.7 mm Hg), with the P-M mode in the middle. A significant difference was found among the IOP readings using the NT-4000 (in various modes), NT-2000, and Goldmann tonometry (repeated-measures analysis of variance, p < 0.01). More than 80% of the results from the NT-4000 were within 3 mm Hg of those from the Goldmann tonometry. Conclusions. The pulse detection feature from this new NCT could capture and monitor pulse waves, thus reducing the variations in the consecutive measurements. The NT-4000 also performed better than the NT-2000 when compared with the Goldmann tonometry.
引用
收藏
页码:601 / 605
页数:5
相关论文
共 24 条
[1]  
American Optometric Association Consensus Panel on Care of the Patient with Age Related Macular Degeneration, 2004, OPT CLIN PRACT GUID
[2]   Causes of blindness and visual impairment in the West of Scotland [J].
Bamashmus, MA ;
Matlhaga, B ;
Dutton, GN .
EYE, 2004, 18 (03) :257-261
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Comparison of the performance of the Nidek NT-2000 noncontact tonometer with the Keeler Pulsair 2000 and the Goldmann applanation tonometer [J].
Cho, P ;
Lui, T .
OPTOMETRY AND VISION SCIENCE, 1997, 74 (01) :51-58
[5]  
Congdon N, 2004, ARCH OPHTHALMOL-CHIC, V122, P477
[6]   Glaucoma in China: how big is the problem? [J].
Foster, PJ ;
Johnson, GJ .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2001, 85 (11) :1277-1282
[7]   The effect of axial length on ocular blood flow assessment in anisometropes [J].
Lam, AKC ;
Chan, ST ;
Chan, B ;
Chan, H .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2003, 23 (04) :315-320
[8]   The effect of myopic axial elongation and posture on the pulsatile ocular blood flow in young normal subjects [J].
Lam, AKC ;
Wong, S ;
Lam, CSY ;
To, CH .
OPTOMETRY AND VISION SCIENCE, 2002, 79 (05) :300-305
[9]   A preliminary study on the Ocular Blood Flow (OBF) of Hong Kong Chinese [J].
Lam, AKC ;
Chan, H ;
Fan, W ;
To, CH .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 1999, 19 (06) :512-517
[10]   Collaborative initial glaucoma treatment study comparing initial treatment randomized to medications or surgery [J].
Lichter, PR ;
Musch, DC ;
Gillespie, BW ;
Guire, KE ;
Janz, NK ;
Wren, PA ;
Mils, RP .
OPHTHALMOLOGY, 2001, 108 (11) :1943-1953