Technical accuracy of a neuronavigation system measured with a high-precision mechanical micromanipulator

被引:58
作者
Kaus, M
Steinmeier, R
Sporer, T
Ganslandt, O
Fahlbusch, R
机构
[1] UNIV ERLANGEN NURNBERG, NEUROCHIRURG KLIN, D-91054 ERLANGEN, GERMANY
[2] UNIV ERLANGEN NURNBERG, LEHRSTUHL TECH ELEKT, D-91054 ERLANGEN, GERMANY
关键词
accuracy; micromanipulator; neuronavigation;
D O I
10.1097/00006123-199712000-00046
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: This study was designed to determine and evaluate the different system-inherent sources of erroneous target localization of a light-emitting diode (LED)-based neuronavigation system (StealthStatisn, Stealth Technologies, Boulder, CO). METHODS: The localization accuracy was estimated by applying a high-precision mechanical micromanipulator to move and exactly locate (+/-0.1 mu m) the pointer at multiple positions in the physical three-dimensional space. The localization error was evaluated by calculating the spatial distance between the (known) LED positions and the LED coordinates measured by the neuronavigator, The results are based on a study of approximately 280,000 independent coordinate measurements. RESULTS: The maximum localization error detected was 0.55 +/- 0.29 mm, with the z direction (distance to the camera array) being the most erroneous coordinate. Minimum localization error was found at a distance of 1400 mm from the central camera (optimal measurement position). Additional error due to 1) mechanical vibrations of the camera tripod (+/-0.15 mm) and the reference frame (+/-0.08 mm) and 2) extrapolation of the pointer tip position from the LED coordinates of at least +/-0.12 mm were detected, leading to a total technical error of 0.55 +/- 0.64 mm. CONCLUSIONS: Based on this technical accuracy analysis, a set of handling recommendations is proposed, leading to an improved localization accuracy. The localization error could be reduced by 0.3 +/- 0.15 mm by correct camera positioning (1400 mm distance) plus 0.15 mm by vibration-eliminating fixation of the camera. Correct handling of the probe during the operation may improve the accuracy by up to 0.1 mm.
引用
收藏
页码:1431 / 1436
页数:6
相关论文
共 14 条
[1]  
ADAMS L, 1990, NATO ADV SCI I F-COM, V60, P411
[2]  
[Anonymous], 1990, PROBABILITY STAT MOD
[3]  
BUCHHOLZ RD, 1992, INTERACTIVE IMAGE GU, P179
[4]  
DACRUZ F, 1987, KERMIT FILE TRANSFER
[5]  
Etter D.M., 1993, ENG PROBLEM SOLVING
[6]  
KAUS M, 1996, GENAUIGKEITSANALYSEN
[7]   THE APPLICATION ACCURACY OF STEREOTAXIC FRAMES [J].
MACIUNAS, RJ ;
GALLOWAY, RL ;
LATIMER, JW .
NEUROSURGERY, 1994, 35 (04) :682-694
[8]  
Press W. H., 1992, Numerical recipes in C++: The Art of Scientific Computing, V2nd
[9]   CORRECTION OF SPATIAL DISTORTION IN MR IMAGING - A PREREQUISITE FOR ACCURATE STEREOTAXY [J].
SCHAD, L ;
LOTT, S ;
SCHMITT, F ;
STURM, V ;
LORENZ, WJ .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1987, 11 (03) :499-505
[10]   In vivo accuracy testing and clinical experience with the ISG viewing wand [J].
Sipos, EP ;
Tebo, SA ;
Zinreich, SJ ;
Long, DM ;
Brem, H .
NEUROSURGERY, 1996, 39 (01) :194-202