Finite element analyses of uniform current density electrodes for radio-frequency cardiac ablation

被引:66
作者
Tungjitkusolmun, S
Woo, EJ
Cao, H
Tsai, JZ
Vorperian, VR
Webster, JG
机构
[1] Univ Wisconsin, Dept Biomed Engn, Madison, WI 53706 USA
[2] Univ Wisconsin, Dept Elect & Comp Engn, Madison, WI 53706 USA
[3] Kyung Hee Univ, Sch Elect & Informat, Yongin 449701, Kyongki Do, South Korea
[4] Univ Wisconsin, Dept Med, Madison, WI 53792 USA
关键词
ablation; bioheat equation; cardiac ablation; current density; electrode; finite element analysis; radio-frequency (RF) ablation; RF ablation; uniform current density;
D O I
10.1109/10.817617
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The high current density at the edge of a metal electrode causes hot spots, which can lead to charring or blood coagulation formation during radio-frequency (RF) cardiac ablation, We used finite element analysis to predict the current density distribution created by several electrode designs for RF ablation, The numerical results demonstrated that there were hot spots at the edge of the conventional tip electrode and the insulating catheter, By modifying the shape of the edge of the 5-mm tip electrode, we could significantly reduce the high current density at the electrode-insulator interface. We also studied the current density distribution produced by a cylindrically shaped electrode. We modified the shape of a cylindrical electrode by recessing the edge and filled in a coating material so that the overall structure was still cylindrical. We analyzed the effects of depth of recess and the electrical conductivity of the added material. The results show that more uniform current density can be accomplished by recessing the electrode, adding a curvature to the electrode, and by coating the electrode with a resistive material.
引用
收藏
页码:32 / 40
页数:9
相关论文
共 25 条
[1]  
[Anonymous], 1985, HEAT TRANSFER MED BI
[2]   RECENT DEVELOPMENTS IN MODELING HEAT-TRANSFER IN BLOOD-PERFUSED TISSUES [J].
ARKIN, H ;
XU, LX ;
HOLMES, KR .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (02) :97-107
[3]   RADIOFREQUENCY CATHETER ABLATION OF THE ATRIA REDUCES INDUCIBILITY AND DURATION OF ATRIAL-FIBRILLATION IN DOGS [J].
ELVAN, A ;
PRIDE, HP ;
EBLE, JN ;
ZIPES, DP .
CIRCULATION, 1995, 91 (08) :2235-2244
[4]  
FOSTER KR, 1989, CRIT REV BIOMED ENG, V17, P25
[5]  
GEUZE RH, 1983, MED BIOL ENG COMPUT, V21, P518, DOI 10.1007/BF02442644
[6]  
Haines DE, 1997, P ANN INT IEEE EMBS, V19, P2628, DOI 10.1109/IEMBS.1997.756871
[7]  
HUANG SKS, 1995, RADIOFREQUENCY CATHE
[8]  
Jain R K, 1980, Ann N Y Acad Sci, V335, P48, DOI 10.1111/j.1749-6632.1980.tb50736.x
[9]  
JALIFE J, 1995, CARDIAC ELECTROPHYSI
[10]   A MINIMUM PROFILE UNIFORM CURRENT-DENSITY ELECTRODE [J].
KSIENSKI, DA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (07) :682-692