Pre-metrological (pre-analytical) variation of some biochemical quantities

被引:17
作者
Fuentes-Arderiu, X [1 ]
Acebes-Frieyro, G [1 ]
Gavaso-Navarro, L [1 ]
Castiñeiras-Lacambra, MJ [1 ]
机构
[1] Ciutat Sanitaria & Univ Bellvitge, Serv Bioquim Clin, Lhospitalet De Llobregat 08907, Catalonia, Spain
关键词
pre-metrological variation; pre-analytical variation; uncertainty of measurement;
D O I
10.1515/CCLM.1999.146
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
It is customary to assume that pre-metrological variation is negligible if all pre-metrological sources of variation have been standardised. The aim of this study was to quantify the pre-metrological variation of some biochemical quantities frequently measured in clinical laboratories. The experimental design considers the different sources of pre-metrological variation that usually act day-to-day on samples from patients. The results demonstrate that, for many quantities, the pre-metrological variance is not negligible; consequently, it is suggested that these variances should be taken into account when estimating the uncertainty of measurement of patients' results.
引用
收藏
页码:987 / 989
页数:3
相关论文
共 6 条
[1]  
BOKELUND H, 1974, CLIN CHEM, V20, P1507
[2]  
FRASER CG, 1989, CRIT REV CL LAB SCI, V27, P409
[3]  
GUDER WG, 1996, PATIENT LAB IMPACT P
[4]  
*INT ORG STAND, 1993, QUANT UN ISO STAND H
[5]  
*INT UN PUR APPL C, 1995, COMP TERM NOM PROP C
[6]  
Young DS, 1997, EFFECTS PREANALYTICA