Nationwide evaluation of X-ray trends survey of abdomen and lumbosacral spine radiography

被引:12
作者
Spelic, DC
Kaczmarek, RV
Suleiman, OH
机构
[1] US FDA, Ctr Devices & Radiol Hlth, Div Mammog Qual, Rockville, MD 20850 USA
[2] US FDA, Radiat Programs, Rockville, MD 20850 USA
关键词
abdomen; radiography; dosimetry; radiations; exposure to patients and personnel; measurement; screens and films; special reports; spine; radiograph;
D O I
10.1148/radiol.2321020397
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Results of the 1995 Nationwide Evaluation of X-ray Trends (NEXT) survey of facilities that perform diagnostic radiographic examinations of the abdomen and lumbosacral spine were compared with those of previous NEXT surveys conducted in 1987 and 1989. A clinically validated radiographic phantom was used in the 1995 survey to capture data about radiation exposure and image quality. Additional data were obtained regarding clinical techniques, facility workloads, x-ray beam quality, film processing quality, and darkroom fog. Mean skin-entrance. air kerma for the abdomen examination dropped from 3.2 mGy (in 1987) to 2.8 mGy at hospitals and from 3.4 mGy (in 1989) to 3.0 mGy at nonhospital facilities. Mean skin-entrance air kerma also decreased for the lumbosacral spine examination from 3.7 mGy (in 1987) to 3.3 mGy at hospitals and from 3.8 mGy (in 1989) to 3.2 mGy at nonhospital facilities. The quality of film processing improved, although 58 (18.3%) of 317 surveyed facilities did not meet the Mammography Quality Standards Act standard for film processing quality, compared with 185 (5.9%) of 3,120 mammography facilities inspected in 1995. Finally, 181 (58.0%) of 312 surveyed facilities had darkroom fog levels greater than the Mammography Quality Standards Act standard, compared with 1,426 (16.6%) of 8,605 mammography facilities inspected in 1995.
引用
收藏
页码:115 / 125
页数:11
相关论文
共 20 条
[1]  
*AM HOSP ASS, 1993, AHA GUID HLTH CAR FI
[2]  
*C RAD CONTR PROGR, 1991, C RAD CONTR PROGR DI, V903
[3]  
*C RAD CONTR PROGR, 1996, C RAD CONTR PROGR DI, V963
[4]  
*C RAD CONTR PROGR, 1998, C RAD CONTR PROGR DI, V982
[5]  
*C RAD CONTR PROGR, 1994, C RAD CONTR PROGR DI, V942
[6]  
*C RAD CONTR PROGR, 2000, C RAD CONTR PROGR DI, V2
[7]  
*C RAD CONTR PROGR, 1989, C RAD CONTR PROGR DI, V893
[8]   NATIONAL SURVEY OF MAMMOGRAPHIC FACILITIES IN 1985, 1988, AND 1992 [J].
CONWAY, BJ ;
SULEIMAN, OH ;
RUETER, FG ;
ANTONSEN, RG ;
SLAYTON, RJ .
RADIOLOGY, 1994, 191 (02) :323-330
[9]   AVERAGE RADIATION-DOSE IN STANDARD CT EXAMINATIONS OF THE HEAD - RESULTS OF THE 1990 NEXT SURVEY [J].
CONWAY, BJ ;
MCCROHAN, JL ;
ANTONSEN, RG ;
RUETER, FG ;
SLAYTON, RJ ;
SULEIMAN, OH .
RADIOLOGY, 1992, 184 (01) :135-140
[10]   A PATIENT-EQUIVALENT ATTENUATION PHANTOM FOR ESTIMATING PATIENT EXPOSURES FROM AUTOMATIC EXPOSURE CONTROLLED X-RAY EXAMINATIONS OF THE ABDOMEN AND LUMBOSACRAL SPINE [J].
CONWAY, BJ ;
DUFF, JE ;
FEWELL, TR ;
JENNINGS, RJ ;
ROTHENBERG, LN ;
FLEISCHMAN, RC .
MEDICAL PHYSICS, 1990, 17 (03) :448-453