Fixation of the tibial component using CMW-1 or Palacos bone cement with gentamicin -: Similar outcome in a randomized radiostereometric study of 51 total knee arthroplasties

被引:21
作者
Adalberth, G [1 ]
Nilsson, KG
Kärrholm, J
Hassander, H
机构
[1] Uppsala Univ, Dept Orthopaed, SE-75185 Uppsala, Sweden
[2] Uppsala Univ, Dept Chem Engn, SE-75185 Uppsala, Sweden
[3] Uppsala Univ, Dept Orthopaed, SE-75185 Umea, Sweden
[4] Uppsala Univ, Dept Orthopaed, SE-75185 Gothenburg, Sweden
来源
ACTA ORTHOPAEDICA SCANDINAVICA | 2002年 / 73卷 / 05期
关键词
D O I
10.1080/000164702321022802
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We studied CMW-1 bone cement with gentamicin in the laboratory and in a randomized clinical study. Palacos bone cement containing gentamicin was used as the control. In the preclinical evaluation, the CMW cement had slightly less mechanical strength. In the clinical study, 51 patients (51 knees) operated on with total knee arthroplasty were studied for 2 years. We used radiostereometric analysis to measure migration of the tibial components, randomized to fixation with either of the two types of cement. The extent and pattern of migration were similar in both groups, and we found no differences in the number, size and extent of radiolucent lines or clinical outcome. No complications occurred. Our findings suggest a need for more studies of CMW-1 bone cement containing gentamicin in a larger cohort of patients.
引用
收藏
页码:531 / 538
页数:8
相关论文
共 27 条
[1]   Low-conforming all-polyethylene tibial component not inferior to metal-backed component in cemented total knee arthroplasty -: Prospective, randomized radiostereometric analysis study of the AGC total knee prosthesis [J].
Adalberth, G ;
Nilsson, KG ;
Byström, S ;
Kolstad, K ;
Milbrink, J .
JOURNAL OF ARTHROPLASTY, 2000, 15 (06) :783-792
[2]  
Adalberth G, 1999, Am J Knee Surg, V12, P233
[3]  
Ahlback S., 1968, Acta Radiol Diagn (Stockh), V277, P7
[4]  
EWALD FC, 1989, CLIN ORTHOP RELAT R, P9
[5]  
Freeman MAR, 1999, J ARTHROPLASTY, V14, P1
[6]   THE EFFECT OF THE TYPE OF CEMENT ON EARLY REVISION OF CHARNLEY TOTAL HIP PROSTHESES - A REVIEW OF 8,579 PRIMARY ARTHROPLASTIES FROM THE NORWEGIAN ARTHROPLASTY REGISTER [J].
HAVELIN, LI ;
ESPEHAUG, B ;
VOLLSET, SE ;
ENGESAETER, LB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77 (10) :1543-1550
[7]  
INSALL JN, 1989, CLIN ORTHOP RELAT R, P13
[8]   MIGRATION OF POROUS COATED ACETABULAR PROSTHESES FIXED WITH SCREWS - ROENTGEN STEREOPHOTOGRAMMETRIC ANALYSIS [J].
KARRHOLM, J ;
SNORRASON, F .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1992, 10 (06) :826-835
[9]   INNOVATIONS IN ACRYLIC BONE-CEMENT AND APPLICATION EQUIPMENT [J].
KINDTLARSEN, T ;
SMITH, DB ;
JENSEN, JS .
JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) :75-83
[10]  
Kuhn KD., 2000, BONE CEMENTS DATE CO