Endovascular aneurysm repair in high-risk patients

被引:108
作者
Chuter, TAM
Reilly, LM
Faruqi, RM
Kerlan, RB
Sawhney, R
Canto, CJ
LaBerge, JM
Wilson, MW
Gordon, RL
Wall, SD
Rapp, J
Messina, LM
机构
[1] Univ Calif San Francisco, Div Vasc Surg, San Francisco, CA 94143 USA
[2] Univ Calif San Francisco, Div Intervent Radiol, San Francisco, CA 94143 USA
关键词
D O I
10.1016/S0741-5214(00)70074-7
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: The purpose of this study was to evaluate the role of endovascular aneurysm repair in high-risk patients. Methods: The elective endovascular repair of infrarenal aortic aneurysm was performed in 116 high-risk patients with either custom-made or commercial stent grafts. The routine follow-up examination included contrast-enhanced computed tomography (CT) before discharge, at 3, 6, and 12 months, and annually thereafter. Patients with endoleak on the initial CT underwent re-evaluation at 2 weeks. Those patients with positive CT results at 2 weeks underwent endovascular treatment. Results: Endovascular repair was considered feasible in 67% of the patients. The mean age was 75 years, and the mean aneurysm diameter was 6.3 cm, The American Society of Anesthesiologists grade was II in 3.4%, III in 65.5%, IV in 30.1%, and V in 0.9%. There were no conversions to open repair. Custom-made aortomonoiliac stent grafts were implanted in 77.6% of the cases, custom-made aortoaotic stent grafts in 11.2%, and commercial bifurcated stent grafts in 11.2%. The 30-day rates of mortality, major morbidity, and minor morbidity were 3.4%, 20.7%, and 12%, respectively, in the first 58 patients and 0%, 3.4%, and 3.4%, respectively in the last 58. The late complications included five cases of stent graft kinking, two cases of femorofemoral graft occlusion, and three cases of proximal stent migration, one of which led to aneurysm rupture. At 2 weeks after repair, endoleak was present in 10.3% of the cases, All the type I (direct perigraft) endoleaks underwent successful endovascular treatment, whereas only one type II (collateral) endoleak responded to treatment. The technical success rare at 2 weeks was 86.2%, and the clinical success rate was 96.6%. The continuing success rate was 87.9%. Seventeen patients died late, unrelated deaths. Conclusion: Endovascular aneurysm repair is safe and effective in patients at high risk, for whom it may be the preferred method of treatment.
引用
收藏
页码:122 / 132
页数:11
相关论文
共 37 条
  • [1] Reporting standards for infrarenal endovascular abdominal aortic aneurysm repair
    Ahn, SS
    Rutherford, RB
    Johnston, KW
    May, J
    Veith, FJ
    Baker, JD
    Ernst, CB
    Moore, WS
    [J]. JOURNAL OF VASCULAR SURGERY, 1997, 25 (02) : 405 - 410
  • [2] Thrombus distribution and changes in aneurysm size following endovascular aortic aneurysm repair
    Armon, MP
    Yusuf, SW
    Whitaker, SC
    Gregson, RHS
    Wenham, PW
    Hopkinson, BR
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1998, 16 (06) : 472 - 476
  • [3] Boyle JR, 1997, J ENDOVASC SURG, V4, P62, DOI 10.1583/1074-6218(1997)004<0062:IRFAAC>2.0.CO
  • [4] 2
  • [5] Abdominal aortic aneurysm in high-risk patients: Short- to intermediate-term results of endovascular repair
    Chuter, TAM
    Gordon, RL
    Reilly, LM
    Kerlan, RK
    Sawhney, R
    Jean-Claude, J
    Canto, CJ
    LaBerge, JM
    Ring, EJ
    Wall, SD
    Messina, LM
    [J]. RADIOLOGY, 1999, 210 (02) : 361 - 365
  • [6] Chuter TAM, 1997, J ENDOVASC SURG, V4, P307, DOI 10.1583/1074-6218(1997)004<0307:SROTIA>2.0.CO
  • [7] 2
  • [8] Chuter TAM, 1998, J ENDOVASC SURG, V5, P259, DOI 10.1583/1074-6218(1998)005<0259:FAEFEA>2.0.CO
  • [9] 2
  • [10] FEMOROFEMORAL BYPASS GRAFT - ANALYSIS OF PATENCY AND FACTORS INFLUENCING LONG-TERM OUTCOME
    CRIADO, E
    BURNHAM, SJ
    TINSLEY, EA
    JOHNSON, G
    KEAGY, BA
    KERSTEIN, MD
    ABURAHMA, AF
    [J]. JOURNAL OF VASCULAR SURGERY, 1993, 18 (03) : 495 - 505