Scatter correction method for x-ray CT using primary modulation: Phantom studies

被引:77
作者
Gao, Hewei [1 ]
Fahrig, Rebecca [1 ]
Bennett, N. Robert [2 ]
Sun, Mingshan [3 ]
Star-Lack, Josh [3 ]
Zhu, Lei [4 ,5 ]
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[2] Qualcomm MEMS Technol, San Jose, CA 95134 USA
[3] Ginzton Technol Ctr, Mountain View, CA 94043 USA
[4] Georgia Inst Technol, Nucl & Radiol Engn Program, George W Woodruff Sch Mech Engn, Atlanta, GA 30332 USA
[5] Georgia Inst Technol, Med Phys Program, George W Woodruff Sch Mech Engn, Atlanta, GA 30332 USA
关键词
aluminium; biomedical materials; computerised tomography; copper; phantoms; X-ray imaging; X-ray scattering; CONE-BEAM CT; CORRECTION ALGORITHM; DIGITAL RADIOGRAPHY; SLOT-SCAN; AIR GAPS; GLARE; SUPPRESSION; REDUCTION; MODEL; RECONSTRUCTION;
D O I
10.1118/1.3298014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Methods: The authors first analyze the potential errors of the estimated scatter in the primary modulation method. On two tabletop CT systems, the method is investigated using three phantoms: A Catphan(C)600 phantom, an anthropomorphic chest phantom, and the Catphan(C)600 phantom with two annuli. Two different primary modulators are also designed to show the impact of the modulator parameters on the scatter correction efficiency. The first is an aluminum modulator with a weak modulation and a low modulation frequency, and the second is a copper modulator with a strong modulation and a high modulation frequency. Results: On the Catphan(C)600 phantom in the first study, the method reduces the error of the CT number in the selected regions of interest (ROIs) from 371.4 to 21.9 Hounsfield units (HU); the contrast to noise ratio also increases from 10.9 to 19.2. On the anthropomorphic chest phantom in the second study, which represents a more difficult case due to the high scatter signals and object heterogeneity, the method reduces the error of the CT number from 327 to 19 HU in the selected ROIs and from 31.4% to 5.7% on the overall average. The third study is to investigate the impact of object size on the efficiency of our method. The scatter-to-primary ratio estimation error on the Catphan(C)600 phantom without any annulus (20 cm in diameter) is at the level of 0.04, it rises to 0.07 and 0.1 on the phantom with an elliptical annulus (30 cm in the minor axis and 38 cm in the major axis) and with a circular annulus (38 cm in diameter). Conclusions: On the three phantom studies, good scatter correction performance of the proposed method has been demonstrated using both image comparisons and quantitative analysis. The theory and experiments demonstrate that a strong primary modulation that possesses a low transmission factor and a high modulation frequency is preferred for high scatter correction accuracy.
引用
收藏
页码:934 / 946
页数:13
相关论文
共 38 条
[1]
Cone beam X-ray scatter removal via image frequency modulation and filtering [J].
Bani-Hashemi, A ;
Blanz, E ;
Maltz, J ;
Hristov, D ;
Svatos, M .
MEDICAL PHYSICS, 2005, 32 (06) :2093-2093
[2]
Potential of software-based scatter corrections in cone-beam volume CT [J].
Bertram, M ;
Wiegert, J ;
Rose, G .
MEDICAL IMAGING 2005: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2005, 5745 :259-270
[3]
Grid and slot scan scatter reduction in mammography: Comparison by using Monte Carlo techniques [J].
Boone, JM ;
Seibert, JA ;
Tang, CM ;
Lane, SM .
RADIOLOGY, 2002, 222 (02) :519-527
[5]
AN ANALYTICAL MODEL OF THE SCATTERED RADIATION DISTRIBUTION IN DIAGNOSTIC-RADIOLOGY [J].
BOONE, JM ;
SEIBERT, JA .
MEDICAL PHYSICS, 1988, 15 (05) :721-725
[6]
Simplified method of scatter correction using a beam-stop-array algorithm for cone-beam computed tomography breast imaging [J].
Cai, Weixing ;
Ning, Ruola ;
Conover, David .
OPTICAL ENGINEERING, 2008, 47 (09)
[7]
The effect of scatter and glare on image quality in contrast-enhanced breast imaging using an a-Si/CsI(Tl) full-field flat panel detector [J].
Carton, Ann-Katherine ;
Acciavatti, Raymond ;
Kuo, Johnny ;
Maidment, Andrew D. A. .
MEDICAL PHYSICS, 2009, 36 (03) :920-928
[8]
Regularization method for scatter-glare correction in fluoroscopic images [J].
Close, RA ;
Shah, KC ;
Whiting, JS .
MEDICAL PHYSICS, 1999, 26 (09) :1794-1801
[9]
PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[10]
SCATTER COMPENSATION FOR DIGITAL CHEST RADIOGRAPHY USING MAXIMUM-LIKELIHOOD EXPECTATION MAXIMIZATION [J].
FLOYD, CE ;
BAYDUSH, AH ;
LO, JY ;
BOWSHER, JE ;
RAVIN, CE .
INVESTIGATIVE RADIOLOGY, 1993, 28 (05) :427-433