Synthesis of an ammonium ionophore and its application in a planar ion-selective electrode

被引:33
作者
Benco, JS [1 ]
Nienaber, HA [1 ]
McGimpsey, WG [1 ]
机构
[1] Worcester Polytech Inst, Dept Chem & Biochem, Worcester, MA 01609 USA
关键词
D O I
10.1021/ac0257851
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A modular technique was used to synthesize an ammonium-selective ionophore based on a cyclic depsipeptide structure. The ionophore was incorporated into a planar ion-selective electrode sensor format and the selectivity tested versus a range of metal cations in a commercial clinical diagnostic "point-of-care" instrument Four sensor membrane formulations were tested, all of which consisted of plasticized PVC. Formulations differed as to the type of plasticizer used and whether an ionic additive was present. It was found that the membrane containing the polar plasticizer nitrophenyl octyl ether in the absence of ionic additive exhibited near-Nernstian behavior (slope, 60.1 mV/decade at 37 degreesC) and possessed high selectivity for ammonium ion over lithium and the divalent cations, calcium and magnesium (log K-NH4+j(POT) = -7.3, -4.4, and -7.1 for lithium, calcium, and magnesium ions, respectively). The same membrane also exhibited sodium and potassium selectivity that was comparable to that reported for nonactin (log K-NH4+j(POT) = -2.1 and -0.6 for sodium and potassium,, respectively, compared to -2.4 and -0.9 in the case of nonactin). Membranes containing the less polar plasticizer, dioctyl phthalate, showed sub-Nernstian behavior (slope, <50 mV/decade at 37 T). In all cases, the presence of the ionic additive potassium tetrakis(4-chlorophenyl)borate substantially reduced the selectivity observed. The flexible modular synthetic technique developed and reported here will allow the cyclic depsipeptide structure to be tuned for optimum selectivity.
引用
收藏
页码:152 / 156
页数:5
相关论文
共 36 条
[1]  
ALEGRET S, 1993, SENSOR ACTUAT B-CHEM, V15, P453
[2]   STUDY OF CALCIUM-ION SELECTIVE ELECTRODES CONTAINING SIMON IONOPHORES USING IMPEDANCE METHODS - BULK PROPERTIES [J].
ARMSTRONG, RD ;
TODD, M .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1988, 257 (1-2) :161-166
[3]   THE ROLE OF PVC IN ION-SELECTIVE ELECTRODE MEMBRANES [J].
ARMSTRONG, RD ;
TODD, M .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1987, 237 (02) :181-185
[4]   Carrier-based ion-selective electrodes and bulk optodes. 1. General characteristics [J].
Bakker, E ;
Buhlmann, P ;
Pretsch, E .
CHEMICAL REVIEWS, 1997, 97 (08) :3083-3132
[5]  
BORCHARDT M, 1995, SENSOR ACTUAT B-CHEM, V24, P721
[6]  
BROWN SS, 1979, CHEM DIAGNOSIS DIS
[7]   RECOMMENDATIONS FOR NOMENCLATURE OF ION-SELECTIVE ELECTRODES - (IUPAC RECOMMENDATIONS 1994) [J].
BUCK, RP ;
LINDNER, E .
PURE AND APPLIED CHEMISTRY, 1994, 66 (12) :2527-2536
[8]   Carrier-based ion-selective electrodes and bulk optodes. 2. Ionophores for potentiometric and optical sensors [J].
Buhlmann, P ;
Pretsch, E ;
Bakker, E .
CHEMICAL REVIEWS, 1998, 98 (04) :1593-1687
[9]   ALL-SOLID-STATE SODIUM-SELECTIVE ELECTRODE BASED ON A CALIXARENE IONOPHORE IN A POLY(VINYL CHLORIDE) MEMBRANE WITH A POLYPYRROLE SOLID CONTACT [J].
CADOGAN, A ;
GAO, ZQ ;
LEWENSTAM, A ;
IVASKA, A ;
DIAMOND, D .
ANALYTICAL CHEMISTRY, 1992, 64 (21) :2496-2501
[10]  
CHAN ADC, 1999, Patent No. 5911862