The investigation of P-32 wire for catheter-based endovascular irradiation

被引:13
作者
Xu, ZG [1 ]
Reinstein, LE [1 ]
Yang, GZ [1 ]
Pai, S [1 ]
Gluckman, G [1 ]
Almond, PR [1 ]
机构
[1] UNIV LOUISVILLE, BROWN CANC CTR, DEPT RADIAT ONCOL, LOUISVILLE, KY 40202 USA
关键词
P-32; endovascular irradiation; restenosis; radiochromic film; brachytherapy dosimetry;
D O I
10.1118/1.597944
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The dose distribution from a P-32 source has been measured and calculated in order to evaluate its application in endovascular irradiation. The source dimension was 27 mm in length and 0.3 mm in diameter and was embedded in the end of a Ni-Ti wire. Dose measurements were performed using radiochromic film in several specially designed tissue equivalent phantoms. Loevinger's point dose kernel was used for the calculation. The approximate dose rate at a radial distance of 1.5 mm from the center of the source was found to be 6.75 cGy/s per GBq (0.25 cGy/s per mCi), which allows the delivery of a therapeutic dose in a short time interval with a satisfactory homogeneity without stepping the source. However, the dose rate falls off almost exponentially along the radial distance. Therefore it may not be suitable for treating large diameter vessel from a centrally located source, The effect of a curved P-32 wire source on the radial dose distribution was also investigated. The results showed that for a maximum bend of 180 degrees the dose rate was increased by as much as 20% along the inner radial distance but decreased by as much as 20% along the outer radial distance compared to the dose along a straight wire. However, for curvatures normally encountered in a clinical situation, the dose rate was changed less than 5%. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:1788 / 1792
页数:5
相关论文
共 23 条
[1]   Dosimetry of a radioactive coronary balloon dilitation catheter for treatment of neointimal hyperplasia [J].
Amols, HI ;
Reinstein, LE ;
Weinberger, J .
MEDICAL PHYSICS, 1996, 23 (10) :1783-1788
[2]  
ARNOLS HI, 1996, INT J RADIAT ONCOL, V36, P913
[3]  
BOTTCHER HD, 1994, INT J RADIAT ONCOL, V29, P183
[4]  
CALIFF RM, 1991, J AM COLL CARDIOL, V17, pB2
[5]   PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY - 10 YEARS EXPERIENCE [J].
CHOKSHI, SK ;
MEYERS, S ;
ABIMANSOUR, P .
PROGRESS IN CARDIOVASCULAR DISEASES, 1987, 30 (03) :147-210
[6]   LOW-DOSE, BETA-PARTICLE EMISSION FROM STENT WIRE RESULTS IN COMPLETE, LOCALIZED INHIBITION OF SMOOTH-MUSCLE CELL-PROLIFERATION [J].
FISCHELL, TA ;
KHARMA, BK ;
FISCHELL, DR ;
LOGES, PG ;
COFFEY, CW ;
DUGGAN, DM ;
NAFTILAN, AJ .
CIRCULATION, 1994, 90 (06) :2956-2963
[7]   DOSE ERRORS DUE TO CHARGE STORAGE IN ELECTRON-IRRADIATED PLASTIC PHANTOMS [J].
GALBRAITH, DM ;
RAWLINSON, JA ;
MUNRO, P .
MEDICAL PHYSICS, 1984, 11 (02) :197-203
[8]   LOW-DOSE RADIOACTIVE ENDOVASCULAR STENTS PREVENT SMOOTH-MUSCLE CELL-PROLIFERATION AND NEOINTIMAL HYPERPLASIA IN RABBITS [J].
HEHRLEIN, C ;
GOLLAN, C ;
DONGES, K ;
METZ, J ;
RIESSEN, R ;
FEHSENFELD, P ;
VONHODENBERG, E ;
KUBLER, W .
CIRCULATION, 1995, 92 (06) :1570-1575
[9]   RESTENOSIS AFTER PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY (PTCA) - A REPORT FROM THE PTCA REGISTRY OF THE NATIONAL-HEART-LUNG-AND-BLOOD-INSTITUTE [J].
HOLMES, DR ;
VLIETSTRA, RE ;
SMITH, HC ;
VETROVEC, GW ;
KENT, KM ;
COWLEY, MJ ;
FAXON, DP ;
GRUENTZIG, AR ;
KELSEY, SF ;
DETRE, KM ;
VANRADEN, MJ ;
MOCK, MB .
AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (12) :C77-C81
[10]   RESTENOSIS AFTER SUCCESSFUL CORONARY ANGIOPLASTY IN PATIENTS WITH SINGLE-VESSEL DISEASE [J].
LEIMGRUBER, PP ;
ROUBIN, GS ;
HOLLMAN, J ;
COTSONIS, GA ;
MEIER, B ;
DOUGLAS, JS ;
KING, SB ;
GRUENTZIG, AR .
CIRCULATION, 1986, 73 (04) :710-717