Pulsatile operation of a centrifugal ventricular assist device with magnetic bearings

被引:23
作者
Bearnson, GB [1 ]
Olsen, DB [1 ]
Khanwilkar, PS [1 ]
Long, JW [1 ]
Allaire, PE [1 ]
Maslen, EH [1 ]
机构
[1] UNIV VIRGINIA,DEPT MECH AEROSP & NUCL ENGN,CHARLOTTESVILLE,VA
关键词
D O I
10.1097/00002480-199609000-00062
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A prototype bench top model of a continuous flow ventricular assist device using an impeller suspended by magnetic bearings has been developed. Generation of a pulsatile pressure was studied using both a computer model and in vitro loop tests of the prototype. The motivation for developing a computer model for a blood pump in the natural circulation is two-fold. First, it allows simulation of the pump under a large variety of operating conditions. Second, it provides insight into what parameters of the system design are important for achieving a specific result. For example, in one case, an aortic pressure of 118/87 mmHg was generated by varying the speed from 2,000 to 2,600 rpm. The computer model was verified by coupling the centrifugal pump prototype to a mock circulatory system. The results of the model were verified by generating an aortic pressure of 113/78 mmHg while varying the speed from 2,000 to 2,600 rpm. These experiments have shown that it is possible to generate pulsatile pressure similar to that of native physiology using a centrifugal left ventricular assist device. further tests will be required to quantify the effects on hemolysis.
引用
收藏
页码:M620 / M624
页数:5
相关论文
共 13 条
[1]  
CRUMP KR, 1990, ASAIO T, V36, P254
[2]  
Joyce L D, 1990, ASAIO Trans, V36, pM619
[3]   BIO-MEDICUS VENTRICULAR ASSIST DEVICE FOR SALVAGE OF CARDIAC SURGICAL PATIENTS [J].
KILLEN, DA ;
PIEHLER, JM ;
BORKON, AM ;
REED, WA .
ANNALS OF THORACIC SURGERY, 1991, 52 (02) :230-235
[4]  
MARCRIS MP, 1994, ASAIO J, V40, pM719
[5]  
PARNIS SM, 1995, ASAIO J, V41, pM333, DOI 10.1097/00002480-199507000-00025
[6]  
RAJAGOPALAN K, 1992, ASAIO ABSTR, V18, P40
[7]  
TATSUMI E, 1995, ASAIO J, V41, pM460, DOI 10.1097/00002480-199507000-00052
[8]  
Tatsumi E, 1992, ASAIO J, V38, pM425, DOI 10.1097/00002480-199207000-00069
[9]  
TODA K, 1995, ASAIO J, V41, pM465, DOI 10.1097/00002480-199507000-00053
[10]  
TOMINAGA T, 1994, ARTIF ORGANS, V18, P252