Impact of prostate volume evaluation by different observers on CT-based post-implant dosimetry

被引:66
作者
Al-Qaisieh, B
Ash, D
Bottomley, DM
Carey, BM
机构
[1] Cookridge Hosp, Dept Med Phys, Leeds LS16 6QB, W Yorkshire, England
[2] Cookridge Hosp, Yorkshire Regional Ctr Canc Treatment, Leeds LS16 6QB, W Yorkshire, England
关键词
prostate brachytherapy; postimplant dosimetry;
D O I
10.1016/S0167-8140(01)00475-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: An analysis Of Computed tomography (CT)-based dosimetry was performed to evaluate the variability of different observers' judgements in marking the prostate gland on CT films, and its effect on the parameters that characterise the prostate implantation quality. Accuracy of data entry by the First author in the process of dosimetry procedure has also been evaluated. Materials and methods: Four observers were asked to evaluate the prostate volume on CT films for six different patients. Each observer repeated the evaluation six times. The sample of patients has a prostate volume in the range of 21.4-42.0 cc derived from transrectal ultrasound volume study. After an average period of 6 weeks of the I-125 implantation, all patients had CT scans. CT-based post-implant dosimetry was performed and the dose volume histograms DVHs were calculated to report the re-constructed prostate Volume, Vp100, Vp150, Vp90 and D90. Comparison between the four observers' output was performed. Results: Comparison between the four observers shows that each observer has a different way of estimating the prostate on CT films. Observers' precision also varies according to the prostate volume and the image quality. This can cause a variation in the resulting D90 value by up to 50%. Analysis of data entry shows a high degree of accuracy. The error of digitizing the prostate is +/-0.19 cc. This is correlated to an error of 0.78 Gy of the D90. Conclusion: The evaluation of prostate gland volume on CT films varies between different observers. This has an effect on the dosimetric indices that characterise the implant quality in particular the D90. (C) 2002 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:267 / 273
页数:7
相关论文
共 9 条
  • [1] ALTMAN GD, 1991, PRACTICAL STAT MED R, P152
  • [2] BLAND M, 1995, INTRO MED STAT, P119
  • [3] Intraobserver and interobserver variability of MR imaging- and CT-derived prostate volumes after transperineal interstitial permanent prostate brachytherapy
    Dubois, DF
    Prestidge, BR
    Hotchkiss, LA
    Prete, JJ
    Bice, WS
    [J]. RADIOLOGY, 1998, 207 (03) : 785 - 789
  • [4] A method for implementing the American Association of Physicists in Medicine Task Group-43 dosimetry recommendations for I-125 transperineal prostate seed implants on commercial treatment planning systems
    Luse, RW
    Blasko, J
    Grimm, P
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 37 (03): : 737 - 741
  • [5] The American Brachytherapy Society recommendations for permanent prostate brachytherapy postimplant dosimetric analysis
    Nag, S
    Bice, W
    DeWyngaery, K
    Prestidge, B
    Stock, R
    Yu, Y
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 46 (01): : 221 - 230
  • [6] NARAYANA V, 1997, INT J RAD ONCOL BIOL, V39, P341
  • [7] DOSIMETRY OF INTERSTITIAL BRACHYTHERAPY SOURCES - RECOMMENDATIONS OF THE AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO 43
    NATH, R
    ANDERSON, LL
    LUXTON, G
    WEAVER, KA
    WILLIAMSON, JF
    MEIGOONI, AS
    [J]. MEDICAL PHYSICS, 1995, 22 (02) : 209 - 234
  • [8] Edema associated with I-125 or Pd-103 prostate brachytherapy and its impact on post-implant dosimetry: An analysis based on serial CT acquisition
    Waterman, FM
    Yue, N
    Corn, BW
    Dicker, AP
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (05): : 1069 - 1077
  • [9] Permanent prostate seed implant brachytherapy: Report of the American Association of Physicists in Medicine Task Group No. 64
    Yu, Y
    Anderson, LL
    Li, ZF
    Mellenberg, DE
    Nath, R
    Schell, MC
    Waterman, FM
    Wu, A
    Blasko, JC
    [J]. MEDICAL PHYSICS, 1999, 26 (10) : 2054 - 2076