A simple theorem relating noise anti patient dose in computed tomography

被引:32
作者
Harpen, MD [1 ]
机构
[1] Univ S Alabama, Dept Radiol, Mobile, AL 36617 USA
关键词
CT; noise; x-ray filters;
D O I
10.1118/1.598778
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Presented is a simple model describing the dependence of image noise in computed tomography on the x-ray beam profile. The model is used to derive the x-ray profile which minimizes total image noise at constant integral patient dose. The profile may be produced with a bow-tie-type beam shaping filter. Results of the analysis are validated using a computer simulation of computed tomography (CT) acquisition and reconstruction. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)01211-0].