Ion trapping at atmospheric pressure (760 Torr) and room temperature with a high-field asymmetric waveform ion mobility spectrometer

被引:44
作者
Guevremont, R [1 ]
Purves, RW
Barnett, DA
Ding, LY
机构
[1] Natl Res Council Canada, Inst Natl Measurement Stand, Ottawa, ON K1A 0R6, Canada
[2] Perkin Elmer Sciex Instruments, Concord, ON L4K 4V8, Canada
[3] Carleton Univ, Ottawa, ON K1S 5B6, Canada
关键词
ion trap; FAIMS; ion mobility;
D O I
10.1016/S1387-3806(99)00148-7
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
A method for the confinement of ions at 760 Torr and room temperature is described. We have recently shown that a cylindrical-geometry high-field asymmetric waveform ion mobility spectrometer (FAIMS), which utilizes an ion separation technique based on the change in ion mobility at high electric fields, focuses ions in two dimensions. This article describes a FAIMS device in which the focusing is extended to three dimensions (i.e. ion trap). Characterization of the ion trap was carried out using a laboratory-constructed time-of-flight mass spectrometer. The half-life of a m/z 380 ion in the trap was determined to be 5 ms. (Int J Mass Spectrom 193 (1999) 45-56) (C) 1999 Elsevier Science B.V.