Diagnostic accuracy of dual-energy CT virtual non-calcium images with different related contrast material values for the detection of bone marrow edema in knee

被引:8
作者
Liang, Jianchao [1 ,2 ]
Fang, Yijie [1 ]
Jiang, Yunping [1 ]
Zhan, Yingying [1 ]
Hong, Guobin [1 ]
机构
[1] Sun Yat Sen Univ, Affiliated Hosp 5, Dept Radiol, Zhuhai 519000, Peoples R China
[2] Zhuhai Peoples Hosp, Dept Radiol, Zhuhai 519000, Peoples R China
关键词
Tomography; X-ray computed; Knee; Bone marrow edema (BME); Dual-energy CT (DECT); Related contrast material; PERFORMANCE; LESIONS;
D O I
10.1016/j.ejrad.2020.109385
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Purpose: The purpose of this study was to evaluate the diagnostic accuracy of different related contrast material (Rel.CM) values in dual-energy computed tomography (DECT) virtual non-calcium (VNCa) images for the detection of bone marrow edema (BME) in knee. Method: This prospective study was approved by the institutional research ethics board, and written informed consent was obtained from all participants. Twenty-three patients (24 knees) who underwent dual-energy CT and MRI within three weeks from July 2018 to June 2019 with a definite history of trauma were enrolled. Each knee was divided into 12 regions. First, MR images served as the reference standard, Receiver operating characteristic (ROC) curve was used and diagnostic accuracy of VNCa images corresponding to different Rel.CM values (1.25, 1.35, 1.45, 1.55, 1.65, 1.75) were analyzed, aimed to select an optimal Rel.CM value of VNCa images for detecting BME. Then, CT values of the normal areas and BME areas were measured on the VNCa images corresponding to the optimal Rel.CM value for preliminary quantitative analysis. The rank-sum test was used to compare the differences of CT values between BME areas and normal bone marrow areas on the VNCa images. Results: The 24 knees were divided into 288 areas. MR Imaging showed BME in 121 areas. The areas under the ROC curve with different Rel.CM values (1.25, 1.35, 1.45, 1.55, 1.65, and 1.75) were 0.633, 0.674, 0.882, 0.684, 0.651, and 0.649, respectively. On the VNCa images of Rel.CM = 1.45, the diagnostic accuracy was the highest (up to 89.2 %), the CT values of the BME area and the normal area were -67.9 (1.7 similar to-100.1) HU and -94.5 (-69.7 similar to-144.9) HU, respectively, with statistical significance (Z=-9.804, P < 0.05). Conclusions: The VNCa images with a Rel.CM value of 1.45 is optimal for the detection of BME in knee.
引用
收藏
页数:6
相关论文
共 20 条
[1]
Akisato K., 2012, ACAD RADIOL, V264, P119, DOI [10.1016/j.acra.2019.02.015, DOI 10.1016/J.ACRA.2019.02.015.]
[2]
Dual-energy CT in vertebral compression fractures: performance of visual and quantitative analysis for bone marrow edema demonstration with comparison to MRI [J].
Bierry, Guillaume ;
Venkatasamy, Aina ;
Kremer, Stephane ;
Dosch, Jean-Claude ;
Dietemann, Jean-Louis .
SKELETAL RADIOLOGY, 2014, 43 (04) :485-492
[3]
Diagnostic accuracy of color-coded virtual noncalcium dual-energy CT for the assessment of bone marrow edema in sacral insufficiency fracture in comparison to MRI [J].
Booz, Christian ;
Noeske, Jochen ;
Albrecht, Moritz H. ;
Lenga, Lukas ;
Martin, Simon S. ;
Bucher, Andreas M. ;
Huizinga, Nicole A. ;
Wichmann, Julian L. ;
Vogl, Thomas J. ;
Yel, Ibrahim .
EUROPEAN JOURNAL OF RADIOLOGY, 2020, 129
[4]
Cao Jianxin, 2014, CHIN J RADIOL, V48, P1013, DOI [10.3760/cma.j.issn.1005-1201.2014.12.012, DOI 10.3760/CMA.J.ISSN.1005-1201.2014.12.012]
[5]
Bone marrow edema in sacroiliitis: detection with dual-energy CT [J].
Chen, Min ;
Herregods, Nele ;
Jaremko, Jacob L. ;
Carron, Philippe ;
Elewaut, Dirk ;
Van den Bosch, Filip ;
Jans, Lennart .
EUROPEAN RADIOLOGY, 2020, 30 (06) :3393-3400
[6]
Bone Marrow Edema at Dual-Energy CT: A Game Changer in the Emergency Department [J].
Gosangi, Babina ;
Mandell, Jacob C. ;
Weaver, Michael J. ;
Uyeda, Jennifer W. ;
Smith, Stacy E. ;
Sodickson, Aaron D. ;
Khurana, Bharti .
RADIOGRAPHICS, 2020, 40 (03) :859-874
[7]
Gu Lingjing, 2014, Zhonghua Yi Xue Za Zhi, V94, P3088
[8]
Diagnostic Performance of Dual-Energy CT for the Detection of Traumatic Bone Marrow Lesions in the Ankle: Comparison with MR Imaging [J].
Guggenberger, Roman ;
Gnannt, Ralph ;
Hodler, Juerg ;
Krauss, Bernhard ;
Wanner, Guido A. ;
Csuka, Esther ;
Payne, Barbara ;
Frauenfelder, Thomas ;
Andreisek, Gustav ;
Alkadhi, Hatem .
RADIOLOGY, 2012, 264 (01) :164-173
[9]
Material differentiation by dual energy CT: initial experience [J].
Johnson, Thorsten R. C. ;
Krauss, Bernhard ;
Sedlmair, Martin ;
Grasruck, Michael ;
Bruder, Herbert ;
Morhard, Dominik ;
Fink, Christian ;
Weckbach, Sabine ;
Lenhard, Miriam ;
Schmidt, Bernhard ;
Flohr, Thomas ;
Reiser, Maximilian F. ;
Becker, Christoph R. .
EUROPEAN RADIOLOGY, 2007, 17 (06) :1510-1517
[10]
Virtual Nonenhanced Dual-Energy CT Urography with Tin-Filter Technology: Determinants of Detection of Urinary Calculi in the Renal Collecting System [J].
Mangold, Stefanie ;
Thomas, Christoph ;
Fenchel, Michael ;
Vuust, Morten ;
Krauss, Bernhard ;
Ketelsen, Dominik ;
Tsiflikas, Ilias ;
Claussen, Claus D. ;
Heuschmid, Martin .
RADIOLOGY, 2012, 264 (01) :119-125